Background
Proximal humerus fractures are often treated with a fixed-angle titanium plate osteosynthesis. Recently, plates made of alternative materials such as carbon fibre-reinforced polyetheretherketone (CFR-PEEK) have been introduced. This study presents the postoperative results of patients treated with a CFR-PEEK plate.
Methods
Patients with proximal humerus fractures treated with a CFR-PEEK plate (PEEKPower™ Humeral Fracture Plate (HFP)) were included. In follow-up examination, age and gender adjusted Constant-Murley Score (ACS), Subjective Shoulder Value (SSV), Quick Disabilities of the Arm, Shoulder and Hand Score (QDASH) and pain score (Visual Analog Scale (VAS)) were analyzed. General condition at follow-up was measured by European Quality of Life 5 Dimensions 3 Level Version (EQ-5D-3L). Range of motion was recorded. In addition, radiographs at follow-up, unfavorable events and revision rate were analyzed.
Results
In total, 98 patients (66.0 ± 13.2 years, 74 females, 24 males) were reexamined. Mean follow-up was 27.6 ± 13.2 months. There were 15 2-part, 28 3-part and 55 4-part fractures. The functional scores showed good results: SSV 83.3 ± 15.6%, QDASH 13.1 ± 17.0 and ACS 80.4 ± 16.0. A 4-part-fracture, head split component, nonanatomic head shaft reposition and preoperative radiological signs of osteoarthritis were significant negative predictors for poorer clinical scores. Unfavourable events were observed in 27 patients (27.6%). Revision surgery was performed in 8 (8.2%) patients. Risk factors for an unfavourable event were female gender, age of 50 years and older, diabetes, affected dominant hand, 4-part fracture, head split and preoperative radiological signs of osteoarthritis.
Conclusion
There are several advantages of the CFR-PEEK plate (PEEKPower™ Humeral Fracture Plate (HFP)) such as the polyaxial screw placement and higher stability of locking screws. In summary, the CFR-PEEK plate osteosynthesis is a good alternative with comparable clinical results and some biomechanical advantages. Proximal humerus fractures show good clinical results after treatment with a CFR-PEEK plate. The revision rate and the risk of unfavorable events are not increased compared to conventional titanium plate osteosynthesis.
Level of evidence
IV
Fracture sequelae type 2 of the proximal humerus-clinical results after arthroplastyLocked dislocations of the glenohumeral joint are rare but often painful and are associated with limited range of motion in the shoulder. In patients of advanced age, arthroplasty is increasingly indicated as a surgical treatment option. Preoperatively, the direction of dislocation, the presence and extent of a glenoid defect, and the soft tissue situation (rotator cuff status, joint capsule) should be analyzed in a differentiated manner. Based on the above factors, we recommend the subclassification of type 2 according to Boileau: posterior locked dislocation (2a), anterior locked dislocation without glenoid defect (2b), and anterior locked dislocation with glenoid defect (2c). In the case of dorsally locked dislocation, a good clinical result can be achieved by using an anatomical endoprosthesis. For ventrally locked dislocations, we recommend using an inverse total endoprosthesis with, if necessary, bony glenoid reconstruction and transfer of the pectoralis major muscle. Level of evidence: IV
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.