The use of the British Journal of Radiology (BJR) (supplement 17) tables of equivalent square fields for dose calculations is widespread. A revised version of the supplement was published recently, with a more elaborate discussion, but without changes in data given in these tables (Br. J. Radiol. suppl 25). The tables were generated for use in dose calculations, with relative beam data such as PDD, BSF, PSF, all with d(max) as the reference depth. However, the current philosophy in dose calculational methods is based on quantities defined at a reference depth, d(ref) = 10 cm, on a separation of phantom and head scatter, and on the use of the relative depth-dose or tissue-phantom ratios normalized at d(ref). By using these quantities as a starting point, problems at shallow depths related to the influence of contaminating electrons in the beam can be eliminated. Recently, a comprehensive set of phantom scatter factor data with d(ref) = 10 cm has been published for a set of square field sizes and a wide range of photon beam energies, showing that phantom scatter is a smoothly varying function of field size and quality index. It is not a priori evident that the conventional concept of equivalent squares for rectangular fields is also fully applicable for phantom scatter factors and phantom scatter related quantities at a depth of 10 cm. It was questioned whether or not new tables of equivalent square fields are needed for this purpose. In this paper, new tables have been constructed for four photon beam energies in the range of Co-60 to 25 MV (quality index from 0.572 to 0.783). The small differences between the outcome of these new tables allowed the construction of one averaged table of equivalent square fields. Phantom scatter factors were calculated for rectangular fields based on the use of the BJR table and on the use of the newly constructed tables and the differences were quantified. For Co-60 no improvements could be shown when using the new averaged table, but for beam energies of 6 to 10 MV small improvements of the order of 0.5 to 1.0% were found. For a higher beam energy of 25 MV the improvement is smaller. Deviations resulting from the BJR table are within the limits of accuracy as stated by the authors. Therefore, for clinical use, the continued use of the BJR table of equivalent squares for phantom scatter factors and phantom scatter related quantities of rectangular fields is justified, irrespective of photon beam energy.
A coherent system for the use of scatter correction factors, determined at 10 cm depth, is described for dose calculations on the central axis of arbitrarily shaped photon beams. The system is suitable for application in both the fixed source-surface distance (SSD) and in the isocentric treatment set-up. This is in contrast to some other proposals where only one of these approaches forms the basis of the calculation system or where distinct quantities and data sets are needed. In order to derive the relations in the formalism, we introduced a separation of the phenomena related to the energy fluence in air and to the phantom scatter contribution to the dose. Both are used relative to quantities defined for the reference irradiation set-up. It is shown that dose calculations can be performed with only one set of basic beam data, obtained at a reference depth of 10 cm. These data consist for each photon beam quality of measured collimator and phantom scatter correction factors, in combination with a set of (percentage/relative) depth-dose or tissue-phantom ratio values measured along the central axis of the beam. Problems related to measurements performed at the depth of maximum absorbed dose, due to the electron contamination of the beam, are avoided in this way. Collimator scatter correction factors are obtained by using a mini-phantom, while phantom scatter correction factors are derived from measurements in a full scatter phantom in combination with the results of the mini-phantom measurements. For practical reasons the fixed SSD system was chosen to determine the data. Then, dose calculations in a fixed SSD treatment set-up itself are straightforward. Application in the isocentric treatment set-up needs simple conversion steps, while the inverse approach, from isocentric to fixed SSD, is described as well. Differences between the two approaches are discussed and the equations for the conversions are given.
The phantom scatter correction factor Sp of megavoltage photon beams can be accurately described using a three-Gaussian fit. The model leads to six parameters, with which Sp(r) is described as a smooth function of the field radius r for beam qualities in the range from 60Co up to 25 MV. The parameters allow Sp values to be calculated at intermediate beam energies and for any field shape. Calculated Sp(X, Y) values for rectangular fields (X, Y) can be subsequently used as reference values to compare with measured Sp(X, Y) values, for example when appraising a new beam.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.