Electrochemical sensor has been fabrication and characterization electrode to detect heavy metals of cadmium with thick film technology. Fabrication electrodes electrochemical sensors start with the design stage of electrodes, screen maker, screen printing, and substrate cutting. The fabrication process produces a ten-pieces of sensors to be used to sensors characterization of SEM, XRD, and Cyclic Voltammetry (CV). The process of characterization produces morphology and the structure of the composites N/IL/G, and linieritas 0,9623, precision ranges between 4,11 and 10,35%, sensitivity of 0,04112 µA.L-1/mg.mm2, limit detection 0,18 mg/L, and accuracy between 79,5 and 126,8%.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.