Laryngeal tubes (LT) are increasingly being used for emergency airway management. This article reports on two patients in whom out-of-hospital intubation with a single-lumen LT was associated with massive pulmonary aspiration in one patient and gastric overinflation in the other. In both cases peak inspiratory pressures exceeded the LT leak pressure of approximately 35 mbar. This resulted in gastric inflation and decreased pulmonary compliance and increased inspiratory pressure further, thereby creating a vicious circle. It is therefore recommended that laryngeal tube suction (LTS) should be used in all cases of emergency airway management and a gastric drain tube be inserted through the dedicated second lumen. Apart from gastric overinflation, incorrect LT/LTS placement must be detected and immediately corrected, e.g. in cases of difficult or impossible gastric tube placement, permanent drainage of air from the gastric tube, decreasing minute ventilation or an ascending capnography curve.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.