We describe a new one-step enzyme immunoassay of troponin T that uses two specific monoclonal antibodies and streptavidin-coated tubes as the solid phase. The monoclonal antibodies were obtained by conventional hybridoma technology, with human troponin T as antigen. The identity of the cardiac troponin T antigen was confirmed by analysis of the amino acid composition and by partial sequence analysis. The specificity of the monoclonal antibodies for cardiac troponin T was proved by immunoblot analysis and displacement curves. The capture antibody is labeled with biotin. The second antibody is conjugated to horseradish peroxidase (EC 1.11.1.7). The assay [Enzymun-TestR system (Boehringer Mannheim GmbH)] is performed in only 90 min at room temperature; the measuring range for troponin T is 0.1 to 15 micrograms/L. The assay shows excellent between-run precision (CV = 3.3-4.9%).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.