The effects of a flexible ventricular restraint device on left ventricular (LV) dilatation and hypertrophy after transmural infarction are examined in an ovine model. Left ventricular remodeling and dilatation occurs after extensive myocardial infarction. A flexible ventricular restraint made from a nitinol mesh was evaluated in adult female sheep (n=14). Cardiac magnetic resonance imaging scans and hemodynamic measurements were completed before and 6 weeks after anterior myocardial infarction. Treatment animals (n=7) received passive ventricular restraint concurrently with LV infarction; the others (n=7) served as controls. Increases in LV end-diastolic volume index were significantly less in the restraint group than in controls (0.20+/-0.41 vs 0.83+/-0.50 ml/kg, p<0.03). End-systolic volumes increased less in treatment animals (0.43+/-0.28 vs 0.90+/-0.38 ml/kg, p<0.03). Control hearts showed an increase in LV mass after infraction, whereas LV mass decreased in restrained hearts (0.14+/-0.19 vs -0.25+/-0.36 g/kg, p<0.03). Hemodynamic studies showed similar changes after infarction for the control and the device group. Gross and microscopic examination showed no device-induced epicardial injury. A flexible ventricular restraint device attenuated remodeling after acute myocardial infarction in sheep.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.