A standard material containing chrysotile asbestos for the validation of x-ray diffractometric quantitation was developed using an asbestos-containing building material i.e., perlite board. The board as the base material was crushed, pulverized, and homogenized. The homogeneity of the powder of perlite board was estimated by analysis of variance. The diffraction intensity values of the crystalline phases and the concentrations of elements determined by x-ray diffractometry and x-ray fluorescence analysis were used for analysis of variance. There is no significant difference between the within-bottle variance and the between-bottle variance, indicating that the powdered perlite board was sufficiently homogenous. The concentration of chrysotile in the material was determined using two methods: an internal standard/x-ray diffractometry method and the x-ray diffractometry/Rietveld refinement. The concentration of chrysotile in the material was determined by an internal standard/x-ray diffractometry method and the material had a chrysotile concentration of 24.1 ± 0.2 mass%.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.