A reciprocal measurement procedure to measure the acoustic feedback path in hearing aids is investigated. The advantage of the reciprocal measurement compared to the direct measurement is a significantly reduced sound pressure in the ear. The direct and reciprocal measurements are compared using measurements on a dummy head with adjustable ear canals, different earmolds, and variations in the outer sound field. The results show that the reciprocal measurement procedure can be used to obtain plausible feedback paths, while reducing the sound pressure in the ear canal by 30 to 40 dB.
The values obtained can be employed to develop a method that represents the shape of the auditory canal as accurately as possible to allow the best possible outcomes for hearing aid fitting.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.