This study aimed to precisely evaluate organ dose and effective dose (E) obtained from a 320-multidetector-row computed tomography (CT) scanner in brain perfusion scans and to estimate the conversion factor (k) between E and dose length product (DLP). A total of 270 thermoluminescent dosemeters were implanted in a male anthropomorphic phantom to measure air kerma. The ratios of mass-energy absorption coefficients were used to convert air kerma into organ doses. The organ doses ranged from 0.01 to 150 mGy. Doses in brain, salivary glands and red bone marrow were relatively high, and dose in eye lens reached about 110 mGy. The resulting effective dose was 5.30 mSv. The resulting conversion factor k = (0.0022 ± 0.0002) mSv·(mGy·cm)−1 was not significantly different from that of 0.0021 mSv·(mGy·cm)−1 reported for head CT scan in ICRP Publication 102.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.