We have developed a simple method for dose calculation in dual asymmetric open and irregular fields with four independent jaws and multileaf collimators. Our calculation method extends the scatter correction method of Kwa et al. [Med. Phys. 21, 1599-1604 (1994)] based on the principle of Day's equivalent-field calculation. The scatter correction factor was determined by the ratio of the derived doses of a smaller asymmetric open field or irregular field to a larger symmetric field. The algorithm with the scatter correction method can be calculated from output factors, tissue maximum ratios, and off-axis ratios for conventional symmetric fields. The doses calculated by this method were compared with the measured doses for various asymmetric open and irregular fields. The agreement between the calculated and measured doses for 4 and 10 MV photon beams was within 0.5% at the geometric center of the asymmetric open fields. For the asymmetric irregular fields with the same geometrical center, agreement within 1% was found in most cases.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.