The dynamic response characteristics of 0.4‐mm‐diameter glass rod thermistors, which are used frequently in fast‐response CTD and microstructure instruments, have been determined over the speed range of 0.08–3.8 m/s. The spectral transfer function was obtained by firing the probes through a thin plume and then comparing the power spectrum of the shot data with that of a slow‐speed map of the plume. The spectral transfer functions are described accurately by a two‐pole filter for frequencies less than 25 Hz, although a single‐pole form is an equally good representation for less than 10 Hz, The frequency of the ½‐power (3‐dB) attenuation point scales as U⅓. Although the spectral forms and velocity scaling of the responses of the different probes tested are the same, the levels show significant differences due to irregularities in the glass coatings around the thermistor material.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.