Radiation exposure to the patient has become a concern for the radiologist in the multidetector computed tomography (CT) era. With the introduction of faster multidetector CT scanners, various techniques have been developed to reduce the radiation dose to the patient; one method is automatic exposure control (AEC). AEC systems make use of different types of control, including patient-size AEC, z-axis AEC, rotational or angular AEC, or a combination of two or more of these types. AEC systems operate on the basis of several methods: standard deviation, noise index, reference milliamperage, and reference image. A clear understanding of how to use different AEC systems on different multidetector CT scanners will allow users to modulate radiation dose, reduce photon starvation artifacts, and maintain image quality throughout the body. Further development of AEC systems and their successful introduction into clinical practice will require user education and good communication between users and manufacturers.
BackgroundTo investigate the sensitivity of various gamma criteria used in the gamma-index method for patient-specific volumetric modulated arc therapy (VMAT) quality assurance (QA) for stereotactic body radiation therapy (SBRT) using a flattening filter free (FFF) photon beam.MethodsThree types of intentional misalignments were introduced to original high-definition multi-leaf collimator (HD-MLC) plans. The first type, referred to Class Out, involved the opening of each bank of leaves. The second type, Class In, involved the closing of each bank of leaves. The third type, Class Shift, involved the shifting of each bank of leaves towards the ground. Patient-specific QAs for the original and the modified plans were performed with MapCHECK2 and EBT2 films. The sensitivity of the gamma-index method using criteria of 1%/1 mm, 1.5%/1.5 mm, 1%/2 mm, 2%/1 mm and 2%/2 mm was investigated with absolute passing rates according to the magnitudes of MLCs misalignments. In addition, the changes in dose-volumetric indicators due to the magnitudes of MLC misalignments were investigated. The correlations between passing rates and the changes in dose-volumetric indicators were also investigated using Spearman’s rank correlation coefficient (γ).ResultsThe criterion of 2%/1 mm was able to detect Class Out and Class In MLC misalignments of 0.5 mm and Class Shift misalignments of 1 mm. The widely adopted clinical criterion of 2%/2 mm was not able to detect 0.5 mm MLC errors of the Class Out or Class In types, and also unable to detect 3 mm Class Shift errors. No correlations were observed between dose-volumetric changes and gamma passing rates (γ < 0.8).ConclusionsGamma criterion of 2%/1 mm was found to be suitable as a tolerance level with passing rates of 90% and 80% for patient-specific VMAT QA for SBRT when using MapCHECK2 and EBT2 film, respectively.
Discrepancies between planned and delivered movements of multi-leaf collimators (MLCs) are an important source of errors in dose distributions during radiotherapy. In this work we used machine learning techniques to train models to predict these discrepancies, assessed the accuracy of the model predictions, and examined the impact these errors have on quality assurance (QA) procedures and dosimetry. Predictive leaf motion parameters for the models were calculated from the plan files, such as leaf position and velocity, whether the leaf was moving towards or away from the isocenter of the MLC, and many others. Differences in positions between synchronized DICOM-RT planning files and DynaLog files reported during QA delivery were used as a target response for training of the models. The final model is capable of predicting MLC positions during delivery to a high degree of accuracy. For moving MLC leaves, predicted positions were shown to be significantly closer to delivered positions than were planned positions. By incorporating predicted positions into dose calculations in the TPS, increases were shown in gamma passing rates against measured dose distributions recorded during QA delivery. For instance, head and neck plans with 1%/2 mm gamma criteria had an average increase in passing rate of 4.17% (SD = 1.54%). This indicates that the inclusion of predictions during dose calculation leads to a more realistic representation of plan delivery. To assess impact on the patient, dose volumetric histograms (DVH) using delivered positions were calculated for comparison with planned and predicted DVHs. In all cases, predicted dose volumetric parameters were in closer agreement to the delivered parameters than were the planned parameters, particularly for organs at risk on the periphery of the treatment area. By incorporating the predicted positions into the TPS, the treatment planner is given a more realistic view of the dose distribution as it will truly be delivered to the patient.
A commercial electron dose calculation software implementation based on the macro Monte Carlo algorithm has recently been introduced. We have evaluated the performance of the system using a standard verification data set comprised of two-dimensional (2D) dose distributions in the transverse plane of a 15 X 15 cm2 field. The standard data set was comprised of measurements performed for combinations of 9-MeV and 20-MeV beam energies and five phantom geometries. The phantom geometries included bone and air heterogeneities, and irregular surface contours. The standard verification data included a subset of the data needed to commission the dose calculation. Additional required data were obtained from a dosimetrically equivalent machine. In addition, we performed 2D dose measurements in a water phantom for the standard field sizes, a 4 cm X 4 cm field, a 3 cm diameter circle, and a 5 cm X 13 cm triangle for the 6-, 9-, 12-, 15-, and 18-MeV energies of a Clinac 21EX. Output factors were also measured. Synthetic CT images and structure contours duplicating the measurement configurations were generated and transferred to the treatment planning system. Calculations for the standard verification data set were performed over the range of each of the algorithm parameters: statistical precision, grid-spacing, and smoothing. Dose difference and distance-to-agreement were computed for the calculation points. We found that the best results were obtained for the highest statistical precision, for the smallest grid spacing, and for smoothed dose distributions. Calculations for the 21EX data were performed using parameters that the evaluation of the standard verification data suggested would produce clinically acceptable results. The dose difference and distance-to-agreement were similar to that observed for the standard verification data set except for the portion of the triangle field narrower than 3 cm for the 6- and 9-MeV electron beams. The output agreed with measurements to within 2%, with the exception of the 3-cm diameter circle and the triangle for 6 MeV, which were within 5%. We conclude that clinically acceptable results may be obtained using a grid spacing that is no larger than approximately one-tenth of the distal falloff distance of the electron depth dose curve (depth from 80% to 20% of the maximum dose) and small relative to the size of heterogeneities. For judicious choices of parameters, dose calculations agree with measurements to better than 3% dose difference and 3-mm distance-to-agreement for fields with dimensions no less than about 3 cm.
The contrast (d = 1) and variance (d = 1) calculated from fluence maps of VMAT plans showed considerable correlations with the plan deliverability, indicating their potential use as indicators for assessing the degree of modulation of VMAT plans. Both contrast and variance consistently showed better performance than the conventional modulation indices for VMAT.
The expanding clinical use of low-energy photon emitting 125I and 103Pd seeds in recent years has led to renewed interest in their dosimetric properties. Numerous papers pointed out that higher accuracy could be obtained in Monte Carlo simulations by utilizing newer libraries for the low-energy photon cross-sections, such as XCOM and EPDL97. The recently developed PENELOPE 2001 Monte Carlo code is user friendly and incorporates photon cross-section data from the EPDL97. The code has been verified for clinical dosimetry of high-energy electron and photon beams, but has not yet been tested at low energies. In the present work, we have benchmarked the PENELOPE code for 10-150 keV photons. We computed radial dose distributions from 0 to 10 cm in water at photon energies of 10-150 keV using both PENELOPE and MCNP4C with either DLC-146 or DLC-200 cross-section libraries, assuming a point source located at the centre of a 30 cm diameter and 20 cm length cylinder. Throughout the energy range of simulated photons (except for 10 keV), PENELOPE agreed within statistical uncertainties (at worst +/- 5%) with MCNP/DLC-146 in the entire region of 1-10 cm and with published EGS4 data up to 5 cm. The dose at 1 cm (or dose rate constant) of PENELOPE agreed with MCNP/DLC-146 and EGS4 data within approximately +/- 2% in the range of 20-150 keV, while MCNP/DLC-200 produced values up to 9% lower in the range of 20-100 keV than PENELOPE or the other codes. However, the differences among the four datasets became negligible above 100 keV.
The Raman intensity for C=C and C≡C peaks increases with an increase in the amount of diacetylene polymerization due to an increase in dose. This study shows the potential of Raman spectroscopy as an alternative for absolute dosimetry verifications with a high-spatial resolution of a few μm, but these findings need to be further validated for the purpose of microdosimetry.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
334 Leonard St
Brooklyn, NY 11211
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.