The current study compared the effectiveness of the graduated tuning fork (128 Hz) and the neurothesiometer in assessing vibration sensation perception in patients presenting with type II diabetes mellitus. A quota sample of patients (n = 21; age range 43-73 years) were assessed using the neurothesiometer and tuning fork by two investigators at five sites on both feet. There was a positive correlation between the results for the two methods of assessment for both investigators, and also between the results for both tools at three individual sites. Overall, there was 66.2% agreement between the results obtained from the two investigators using the tuning fork at each site; however, Kappa values only reached statistical significance at one site, indicating variability between the results from the two tools. This study suggests that assessment of vibration sensation with the tuning fork may be unreliable. These preliminary findings are based on a small sample size; thus further research is warranted.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.