Despite much development, there remains dosimetric uncertainty in the surface and build-up regions in intensity-modulated radiation therapy treatment plans for head and neck cancers. Experiments were performed to determine the dosimetric discrepancies in the surface and build-up region between the treatment planning system (TPS) prediction and experimental measurement using radiochromic film. A head and neck compression film phantom was constructed from two semicylindrical solid water slabs. Treatment plans were generated using two commercial TPSs (PINNACLE3 and CORVUS) for two cases, one with a shallow (approximately 0.5 cm depth) target and another with a deep (approximately 6 cm depth) target. The plans were evaluated for a 54 Gy prescribed dose. For each case, two pieces of radiochromic film were used for dose measurement. A small piece of film strip was placed on the surface and another was inserted within the phantom. Overall, both TPSs showed good agreement with the measurement. For the shallow target case, the dose differences were within +/- 300 cGy (5.6% with respect to the prescribed dose) for PINNACLE3 and +/- 240 cGy (4.4%) for CORVUS in 90% of the region of interest. For the deep target case, the dose differences were +/- 350 (6.5%) for PINNACLE3 and +/- 260 cGy (4.8%) for CORVUS in 90% of the region of interest. However, it was found that there were significant discrepancies from the surface to about 0.2 cm in depth for both the shallow and deep target cases. It was concluded that both TPSs overestimated the surface dose for both shallow and deep target cases. The amount of overestimation ranges from 400 to 1000 cGy (approximately 7.4% to 18.5% with respect to the prescribed dose, 5400 cGy).
The number of imaging data sets has significantly increased during radiation treatment after introducing a diverse range of advanced techniques into the field of radiation oncology. As a consequence, there have been many studies proposing meaningful applications of imaging data set use. These applications commonly require a method to align the data sets at a reference. Deformable image registration (DIR) is a process which satisfies this requirement by locally registering image data sets into a reference image set. DIR identifies the spatial correspondence in order to minimize the differences between two or among multiple sets of images. This article describes clinical applications, validation, and algorithms of DIR techniques. Applications of DIR in radiation treatment include dose accumulation, mathematical modeling, automatic segmentation, and functional imaging. Validation methods discussed are based on anatomical landmarks, physical phantoms, digital phantoms, and per application purpose. DIR algorithms are also briefly reviewed with respect to two algorithmic components: similarity index and deformation models.
A comprehensive set of data on skin dose for 8 MV and 18 MV photon beams from a medical linear accelerator was measured using a parallel-plate chamber to document the effect of field size, source-to-surface distance (SSD), off-axis distance, acrylic block tray, wedge (external standard wedge), Lipowitz's metal block, multileaf collimator (MLC), and dynamic wedge. The skin dose increased as field size increased from 5 X 5 cm2 to 40 X 40 cm2 (6% to 38% for 8 MV and 5% to 44% for 18 MV beam). With the use of an acrylic block tray, the skin dose increased for all field sizes (7% to 59% for 8 MV and 5% to 62% for 18 MV beam), but the increase was minimal for small fields. The skin dose with a wedge showed a much more complex trend. It was generally lower than the dose for an open field, but higher in the case of large fields and higher degree wedges. When both wedge and block tray were used, the tray was a major contributor to the skin dose because some of the contaminant electrons from the wedge assembly were absorbed by the block tray. Field-shaping blocks increased the skin dose, but, interestingly, the block tray reduced the skin dose for small blocked fields treated with a high-energy photon beam. The effect of an MLC on skin dose was very similar to that of a Lipowitz's metal block, but its magnitude was less. The skin dose was higher for dynamic wedge fields than it was for standard wedge fields. As SSD decreased, the skin dose increased, and this effect was dominant in larger field sizes. The SSD effect was enhanced in the presence of an acrylic block tray. The skin dose off-axis was the same as at the central axis, or smaller. A similar pattern of behavior of the skin dose is expected for photon beams from other linear accelerators.
The main purpose of this work was to quantify patient organ doses from the two kilovoltage cone beam computed tomography (CBCT) systems currently available on medical linear accelerators, namely the X‐ray Volumetric Imager (XVI, Elekta Oncology Systems) and the On‐Board Imager (OBI, Varian Medical Systems). Organ dose measurements were performed using a fiber‐optic coupled (FOC) dosimetry system along with an adult male anthropomorphic phantom for three different clinically relevant scan sites: head, chest, and pelvis. The FOC dosimeter was previously characterized at diagnostic energies by Hyer et al. [Med Phys 2009;36(5):1711–16] and a total uncertainty of approximately 4% was found for in‐phantom dose measurements. All scans were performed using current manufacturer‐installed clinical protocols and appropriate bow‐tie filters. A comparison of image quality between these manufacturer‐installed protocols was also performed using a Catphan 440 image quality phantom. Results indicated that for the XVI, the dose to the lens of the eye (1.07 mGy) was highest in a head scan, thyroid dose (19.24 mGy) was highest in a chest scan, and gonad dose (29 mGy) was highest in a pelvis scan. For the OBI, brain dose (3.01 mGy) was highest in a head scan, breast dose (5.34 mGy) was highest in a chest scan, and gonad dose (34.61 mGy) was highest in a pelvis scan. Image quality measurements demonstrated that the OBI provided superior image quality for all protocols, with both better spatial resolution and low‐contrast detectability. The measured organ doses were also used to calculate a reference male effective dose to allow further comparison of the two machines and imaging protocols. The head, chest, and pelvis scans yielded effective doses of 0.04, 7.15, and 3.73 mSv for the XVI, and 0.12, 1.82, and 4.34 mSv for the OBI, respectively.PACS number: 87.57.uq
PurposeThe aim of this study was to investigate the intra‐fractional patient motion using the ExacTrac system in LINAC‐based stereotactic radiosurgery (SRS).MethodA retrospective analysis of 104 SRS patients with kilovoltage image‐guided setup (Brainlab ExacTrac) data was performed. Each patient was imaged pre‐treatment, and at two time points during treatment (1st and 2nd mid‐treatment), and bony anatomy of the skull was used to establish setup error at each time point. The datasets included the translational and rotational setup error, as well as the time period between image acquisitions. After each image acquisition, the patient was repositioned using the calculated shift to correct the setup error. Only translational errors were corrected due to the absence of a 6D treatment table. Setup time and directional shift values were analyzed to determine correlation between shift magnitudes as well as time between acquisitions.ResultsThe average magnitude translation was 0.64 ± 0.59 mm, 0.79 ± 0.45 mm, and 0.65 ± 0.35 mm for the pre‐treatment, 1st mid‐treatment, and 2nd mid‐treatment imaging time points. The average time from pre‐treatment image acquisition to 1st mid‐treatment image acquisition was 7.98 ± 0.45 min, from 1st to 2nd mid‐treatment image was 4.87 ± 1.96 min. The greatest translation was 3.64 mm, occurring in the pre‐treatment image. No patient had a 1st or 2nd mid‐treatment image with greater than 2 mm magnitude shifts.ConclusionThere was no correlation between patient motion over time, in direction or magnitude, and duration of treatment. The imaging frequency could be reduced to decrease imaging dose and treatment time without significant changes in patient position.
Ever since the advent and development of treatment planning systems, the uncertainty associated with calculation grid size has been an issue. Even to this day, with highly sophisticated 3D conformal and intensity-modulated radiation therapy (IMRT) treatment planning systems (TPS), dose uncertainty due to grid size is still a concern. A phantom simulating head and neck treatment was prepared from two semi-cylindrical solid water slabs and a radiochromic film was inserted between the two slabs for measurement. Plans were generated for a 5,400 cGy prescribed dose using Philips Pinnacle(3) TPS for two targets, one shallow ( approximately 0.5 cm depth) and one deep ( approximately 6 cm depth). Calculation grid sizes of 1.5, 2, 3 and 4 mm were considered. Three clinical cases were also evaluated. The dose differences for the varying grid sizes (2 mm, 3 mm and 4 mm from 1.5 mm) in the phantom study were 126 cGy (2.3% of the 5,400 cGy dose prescription), 248.2 cGy (4.6% of the 5,400 cGy dose prescription) and 301.8 cGy (5.6% of the 5,400 cGy dose prescription), respectively for the shallow target case. It was found that the dose could be varied to about 100 cGy (1.9% of the 5,400 cGy dose prescription), 148.9 cGy (2.8% of the 5,400 cGy dose prescription) and 202.9 cGy (3.8% of the 5,400 cGy dose prescription) for 2 mm, 3 mm and 4 mm grid sizes, respectively, simply by shifting the calculation grid origin. Dose difference with a different range of the relative dose gradient was evaluated and we found that the relative dose difference increased with an increase in the range of the relative dose gradient. When comparing varying calculation grid sizes and measurements, the variation of the dose difference histogram was insignificant, but a local effect was observed in the dose difference map. Similar results were observed in the case of the deep target and the three clinical cases also showed results comparable to those from the phantom study.
Purpose: With external beam radiation therapy, uncertainties in treatment planning and delivery can result in an undesirable dose distribution delivered to the patient that can compromise the benefit of treatment. Techniques including geometric margins and probabilistic optimization have been used effectively to mitigate the effects of uncertainties. However, their broad application is inconsistent and can compromise the conclusions derived from cross-technique and cross-modality comparisons. Methods and Materials: Conventional methods to deal with treatment planning and delivery uncertainties are described, and robustness analysis is presented as a framework that is applicable across treatment techniques and modalities. Results: This report identifies elements that are imperative to include when conducting a robustness analysis and describing uncertainties and their dosimetric effects.NotedEarn CME credit by taking a brief online assessment at https://academy.astro.org.Conclusion: The robustness analysis approach described here is presented to promote reliable plan evaluation and dose reporting, particularly during clinical trials conducted across institutions and treatment modalities.
Intrafraction patient motion is much more likely in intensity-modulated radiation therapy (IMRT) than in conventional radiotherapy primarily due to longer beam delivery times in IMRT treatment. In this study, we evaluated the uncertainty of intrafraction patient displacement in CNS and head and neck IMRT patients. Immobilization is performed in three steps: (1) the patient is immobilized with thermoplastic facemask, (2) the patient displacement is monitored using a commercial stereotactic infrared IR camera (ExacTrac, BrainLab) during treatment, and (3) repositioning is carried out as needed. The displacement data were recorded during beam-on time for the entire treatment duration for 5 patients using the camera system. We used the concept of cumulative time versus patient position uncertainty, referred to as an uncertainty time histogram (UTH), to analyze the data. UTH is a plot of the accumulated time during which a patient stays within the corresponding movement uncertainty. The University of Florida immobilization procedure showed an effective immobilization capability for CNS and head and neck IMRT patients by keeping the patient displacement less than 1.5 mm for 95% of treatment time (1.43 mm for 1, and 1.02 mm for 1, and less than 1.0 mm for 3 patients). The maximum displacement was 2.0 mm.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
334 Leonard St
Brooklyn, NY 11211
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.