A rapid method for the quantitative determination of the hydroxyl value (OHV) of hydroxylated soybean oils by HATR/FTIR spectroscopy is described. Calibration standards were prepared by the formic acid/ hydrogen peroxide method and OH values were determined by the official method of AOCS Tx 1a-66, covering an analytical range of 3.5-125 mg of KOH/g of sample. A partial least squares (PLS) calibration model for the prediction of the hydroxyl value (OHV) was developed based on eight different spectral subregions between 3,150 and 990 cm -1 and combinations of them. On average, 36 samples were used for the modeling and 17 were used for external validation. The resulting calibration was linear over the analytical range and had a standard deviation of 2.334. Validation of the method was carried out by comparing the OHV of a series of hydroxylated soybean oils predicted by the PLS model to the values obtained by the AOCS standard method. A correlation coefficient of R 2 = 0.9843 and RMSEC and RMSEP values of, respectively, 3.393 and 3.643 were obtained. After the calibration of the spectrometer, the OHV could be obtained in 2-3 min per sample, a major improvement over conventional wet chemical methods. The advantages of these methodologies are that they do not destroy the sample, have a lower cost, expedite the analysis and do not produce residues. Therefore, they may yield excellent results when used to quantify OHV of soybean polyols obtained by hydroxylation reaction.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.