To determine the precision and agreement of the hemoglobin (Hb) measurements in capillary and venous blood samples by the HemoCue and an automated counter. Hb was determined by both equipaments in blood samples of 29 pregnant women. The HemoCue showed low repeatability of Hb measurements in duplicate in capillary (CR=0.53 g/dL, CV=13.6%) and venous blood (CR=0.53 g/dL, CV=13.6%). Hb measurements in capillary blood were higher than those in venous blood (12.4 and 11.7 g/dL, respectively; p<0.05). There was high agreement between Hb in capillary blood by the HemoCue and in venous blood by the counter (r icc =0.86; p<0.01), and also between the diagnosis of anemia by both equipments (k=0.81; p<0.01). The HemoCue seems to be more appropriate for capillary blood and require training of the measurers. (12,4 e 11,7 g/ dL, respectivamente; p<0,05 =0,86; p<0,01), bem como entre os diagnósticos de anemia realizados nos dois equipamentos (k=0,81; p<0,01). O HemoCue parece ser mais apropriado para análise de Hb em sangue capilar, mas é necessário um treinamento dos analisadores. Resumo
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.