Multidrug resistant tuberculosis and non-tuberculous mycobacterium infections present challenges due to complex treatment regimens. Extended treatment regimes expose patients to higher risks of toxic side-effects. A high drug toxicity profile necessitates closer monitoring. One of the more challenging issues is QTc prolongation with non-injectable regimens.
This study investigates the portable AliveCor device to record and measure the QTc on a 6-lead ECG. An automated QTc readout from 12-Lead ECG for each patient (n = 13) and mean QTc value calculated from each patients’ respective AliveCor tracing were compared. The general trend suggests AliveCor underestimates QTc − 92% cases calculated the AliveCor QTc as lower than their corresponding 12-Lead QTc readout.
The use of AliveCor could potentially be translated into current clinical practice with caution of percentage variation either side. This could facilitate the use of AliveCor as a promising and convenient screening tool before further evaluation by a 12-Lead ECG is required.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.