Single percepts induced by single-electrode stimulation were relatively small, but the form of percepts, especially after multielectrode stimulation, often did not match the stimulation pattern, even in a normal-sighted volunteer. Reproducible percepts were more easily generated than those that matched the stimulation pattern.
The electrical conductivity of human cerebrospinal fluid (CSF) from seven patients was measured at both room temperature (25 degrees C) and body temperature (37 degrees C). Across the frequency range of 10 Hz-10 kHz, room temperature conductivity was 1.45 S/m, but body temperature conductivity was 1.79 S/m, approximately 23% higher. Modelers of electrical sources in the human brain have underestimated human CSF conductivity by as much as 44% for nearly two decades, and this should be corrected to increase the accuracy of source localization models.
Threshold charge densities in severely blind patients were substantially higher than that in a normal-sighted patient. Charge densities in blind patients always exceeded one seemingly reasonable estimate of safe stimulation. The potential adversity of long-term stimulation of the retina by a prosthesis has yet to be determined.
A wirelessly operated, minimally invasive retinal prosthesis was developed for preclinical chronic implantation studies in Yucatan minipig models. The implant conforms to the outer wall of the eye and drives a microfabricated polyimide stimulating electrode array with sputtered iridium oxide electrodes. This array is implanted in the subretinal space using a specially designed ab externo surgical technique that fixes the bulk of the prosthesis to the outer surface of the sclera. The implanted device is fabricated on a host polyimide flexible circuit. It consists of a 15-channel stimulator chip, secondary power and data receiving coils, and discrete power supply components. The completed device is encapsulated in poly(dimethylsiloxane) except for the reference/counter electrode and the thin electrode array. In vitro testing was performed to verify the performance of the system in biological saline using a custom RF transmitter circuit and primary coils. Stimulation patterns as well as pulse strength, duration, and frequency were programmed wirelessly using custom software and a graphical user interface. Wireless operation of the retinal implant has been verified both in vitro and in vivo in three pigs for more than seven months, the latter by measuring stimulus artifacts on the eye surface using contact lens electrodes.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.