To assess the out-of-field surface and internal dose of the 1.5T MR-Linac compared to the conventional external beam linac using optically stimulated luminescence dosimeters (OSLDs), and evaluate the out-of-field dose calculation accuracy of the Monaco treatment planning system of the 1.5T MR-Linac. Methods: A cubic solid water phantom, with OSLDs on the surface, was vertically irradiated by MR-Linac square fields with different sizes. In addition, OSLDs were arranged out of the beam edges in four directions. An anthropomorphic adult phantom, with 125 cm 3 simulated volume, was irradiated in 4 orthogonal directions by both MR-Linac and conventional linac at the head, thoracic, and pelvic sites. Out-of-field doses were measured by OSLDs on both the surface and internal emulational organs at risk (OARs). The results were compared to the simulated dose from Monaco treatment planning system (TPS). Results: At different field sizes (5 × 5 to 20 × 20 cm 2) and distances (1 to 10 cm) to beam edge, the out-of-field surface dose measured on MR-Linac varied from 0.16 % (10 cm to 5 × 5 cm 2 edge) to 7.02 % (1 cm to 20 × 20 cm 2 edge) of the maximum dose laterally and from 0.14 % (10 cm to 5 × 5 cm 2 edge) to 8.56 % (1 cm to 20 × 20 cm 2 edge) of the maximum dose longitudinally. Compared to the OSLDs measured data, the Monaco TPS presented an overestimate of the out-of-field dose of OARs at 0-2 % isodose area on both surface and internal check points, and the overestimation gets greater as the distance increases. The underestimation was found to be 0-35% at 2%-5% Accepted Article This article is protected by copyright. All rights reserved isodose area on both surface and internal check points. Compared to the conventional linac, MR-Linac delivered higher average values of out-of-field dose on surface check points (20 %, 19 %, 21 %) and internal simulated OARs (42 %, 37 %, 9 %) of the anthropomorphic phantom at head, thoracic and pelvic irradiations, respectively. Conclusions: Compared to the conventional linac, MR-Linac has the same out-of-field dose distribution. However, considering the absolute dose values, MR-Linac delivered relatively higher out-of-field doses on both surface and internal OARs. Additional radiation shielding to patients undergoing MR-Linac may provide protection from out-of-field exposure.
Critical biomarkers of disease are increasingly being detected by point‐of‐care assays. Chemiluminescence (CL) and electrochemiluminescence (ECL) are often used in such assays due to their convenience and that they do not require light sources or other components that could complicate or add cost to the system. Reports of these assays often include readers built on a cellphone platform or constructed from low‐cost components. However, the impact the optical design has on the limit of detection (LOD) in these systems remains unexamined. Here, we report a theoretical rubric to evaluate different optical designs in terms of maximizing the use of photons emitted from a CL or ECL assay to improve the LOD. We demonstrate that the majority of cellphone designs reported in the literature are not optimized, in part due to misunderstandings of the optical tradeoffs in collection systems, and in part due to limitations imposed on the designs arising from the use of a mobile phone with a very small lens aperture. Based on the theoretical rubric, we design a new portable reader built using off‐the‐shelf condenser optics, and demonstrate a nearly 10× performance enhancement compared to prior reports on an ECL assays running on a portable chip.
Because of the bulk, complexity, calibration requirements, and need for operator training, most current flow-based blood counting devices are not appropriate for field use. Standard imaging methods could be much more compact, inexpensive, and with minimal calibration requirements. However, due to the diffraction limit, imaging lacks the nanometer precision required to measure red blood cell volumes. To address this challenge, we utilize Mie scattering, which can measure nanometer-scale morphological information from cells, in a dark-field imaging geometry. The approach consists of a custom-built dark-field scattering microscope with symmetrically oblique illumination at a precisely defined angle to record wide-field images of diluted and sphered blood samples. Scattering intensities of each cell under three wavelengths are obtained by segmenting images via digital image processing. These scattering intensities are then used to determine size and hemoglobin information via Mie theory and machine learning. Validation on 90 clinical blood samples confirmed the ability to obtain mean corpuscular volume (MCV), mean corpuscular hemoglobin concentration (MCHC), and red cell distribution width (RDW) with high accuracy. Simulations based on historical data suggest that an instrument with the accuracy achieved in this study could be used for widespread anemia screening.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.