This paper presents the design, fabrication, and characterization of a middle-infrared (MIR) linear variable optical filter (LVOF) and thermopile detectors that will be used in a miniaturized mixed gas detector for CH4/CO2/CO measurement. The LVOF was designed as a tapered-cavity Fabry-Pérot optical filter, which can transform the MIR continuous spectrum into multiple narrow band-pass spectra with peak wavelength in linear variation. Multi-layer dielectric structures were used to fabricate the Bragg reflectors on the both sides of tapered cavity as well as the antireflective film combined with the function of out-of-band rejection. The uncooled thermopile detectors were designed and fabricated as a multiple-thermocouple suspension structure using micro-electro-mechanical system technology. Experimentally, the LVOF exhibits a mean full-width-at-half-maximum of 400 nm and mean peak transmittance of 70% at the wavelength range of 2.3~5 μm. The thermopile detectors exhibit a responsivity of 146 μV/°C at the condition of room temperature. It is demonstrated that the detectors can achieve the quantification and identification of CH4/CO2/CO mixed gas.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.