Aim: The aim of present method development for Diphenhydramine HCl performed on HPLC is to obtain specific, more accurate and precise results as compare to spectrophotometric method. Methods: HPLC analysis was performed according to USP method with wavelength detection at 220nm and 1.0ml/min flow rate. Wufeng thermo HPLC system UV -100detector was used having column C18(4.6mm*250mm) 5. Methanol and water (4:1) mixture was used as mobile phase and pH was adjusted at 7.4 with the help of triethanolamine. Validation parameters like linearity, accuracy, precision, solution stability, robustness, LOD, LOQ and system suitability were successfully evaluated. Results: The regression co-efficient for calibration curve was 0.991 and % recovery was in range (80-110%), whereas no robustness was observed in this reported method. Conclusion: In summary, the expected linearity, accuracy and % recovery indicating that HPLC is more precise method than spectrophotometry and suggested that present method qualifies the validation criteria.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.