Digital radiography is often performed at a higher dose rate than analogue radiography for image acquisition. The authors measured the Entrance Surface Dose (ESD) of analogue and digital radiography techniques for 14 radiographic examinations from randomly selected medical centres in the central district of Korea. It was that the mean ESD of the digital examinations was 2.84 mGy (range, 0.37-6.38 mGy) and that of the analogue examinations was 1.83 mGy (range, 0.38-4.74 mGy), resulting in a 55.25 % higher ESD for digital technique. Although this survey is not completely representative of Korea, findings of this study indicate a need for closer exposure management in digital radiography to minimise patient dose.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.