The aim of the present study was to evaluate and compare the in vitro and flow dynamics of the Magna (MB) and the Magna Ease aortic valve bioprosthesis (MEB) within the ascending aorta. A 2D-particle-image-velocimetry (2D-PIV) study was performed to compare the flow dynamics induced by each pericardial Carpentier-Edwards Magna and Magna Ease aortic valve prosthesis in the aortic flow field directly behind the valve. Both prostheses (diameter 23 mm) were placed inside an artificial aorta under pulsatile flow conditions (70 Hz and 70 ml stroke volume). The flow field was evaluated according to velocity, shear strength, and vorticity. Both prostheses showed a jet flow type profile with a maximum velocity of 0.97±0.09 m/s for MB and 0.83±1.8 m/s for MEB. Flow fields of both valves were similar in acceleration, peak flow deceleration and leakage phase. Maximum shear strength was 20,285±11,774 l/s2 for MB and 17,006±8453 l/s2 for MEB. Vorticity was nearly similar for counterclockwise and clockwise rotation in both prostheses, but slightly higher with MB (251±41 l/s and -250±39 l/s vs. 225±48 l/s and -232±48 l/s). The point-of-interest (POI)-analysis revealed a higher velocity for left-sided aortic wall compared to right-sided at MB (0.12±0.09 m/s vs. 0.18±0.10 m/s, p<0.001), but was consistent at MEB (0.09±0.05 m/s vs. 0.08±0.04 m/s, p=0.508), respectively. Velocity, shear strength and vorticity in an in vitro test set-up are lower with MEB compared to MB, thus resulting in improved flow dynamics with a similar flow field, which might have a positive influence on blood rheology and potential valve degeneration.
The present study clearly confirmed ability of an accelerated cutting of stenotic aortic valve by the aortic valve resection tool. Nearly all leaflets were cut and a small rim was left within the annulus, hence providing an ideal 'landing zone' for the new prosthesis. Nevertheless, the aortic valve resection tool should be enhanced by adding a centering mechanism, thus achieving a more precise cutting process in order to avoid secondary damage.
To date, cardiac valve diseases are considered as a major public health problem and most frequently, the aortic valve is affected. To treat high-risk patients, catheter-based techniques have been developed recently, avoiding open heart surgery and/or cardiopulmonary bypass. Although these sophisticated and rapidly emerging catheter-based technologies do allow a minimally invasive treatment option of high-risk patients on the one hand, further developments and in vitro testing under physiological conditions are necessary, on the other hand, in order to further optimize them for clinical routines. Therefore, we present the concept of a new multifunctional flow channel, offering (i) the possibility of transapical access; (ii) the simulation of physiological flow conditions; and (iii) the evaluation of the fluid flow by 2D particle image velocimetry within a wide range of parameters.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.