Background The accuracy of a new optical biosensor (OB) point-of-care device for the detection of severe infections is studied. Methods The OB emits different wavelengths and outputs information associated with heart rate, pulse oximetry, levels of nitric oxide and kidney function. At the first phase, recordings were done every two hours for three consecutive days after hospital admission in 142 patients at high-risk for sepsis by placing the OB on the forefinger. At the second phase, single recordings were done in 54 patients with symptoms of viral infection; 38 were diagnosed with COVID-19. Results At the first phase, the cutoff value of positive likelihood of 18 provided 100% specificity and 100% positive predictive value for the diagnosis of sepsis. These were 87.5 and 91.7% respectively at the second phase. OB diagnosed severe COVID-19 with 83.3% sensitivity and 87.5% negative predictive value. Conclusions The studied OB seems valuable for the discrimination of infection severity.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citationsācitations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright Ā© 2024 scite LLC. All rights reserved.
Made with š for researchers
Part of the Research Solutions Family.