The GE Discovery MI PET/CT system has a modular digital detector design allowing three, four, or five detector block rings that extend the axial field-of-view (FOV) from 15 to 25 cm in 5 cm increments. This study investigated the performance of the 5-ring system and compared it to 3-and 4-ring systems; the GE Discovery IQ system that uses conventional photomultiplier tubes; and the GE Signa PET/MR system that has a reduced transaxial FOV. Methods: PET performance was evaluated at three different institutions. Spatial resolution, sensitivity, counting rate performance, accuracy, and image quality were measured in accordance with National Electrical Manufacturers Association NU 2-2012 standards. The mean energy resolution, mean timing resolution, and PET/CT subsystem alignment were also measured. Phantoms were used to determine the effects of varying acquisition time and reconstruction parameters on image quality. Retrospective patient scans were reconstructed with various scan durations to evaluate the impact on image quality. Results: Results from all three institutions were similar. Radial/tangential/axial full width at half maximum spatial resolution measurements using the filtered back projection algorithm were 4.3/4.3/ 5.0 mm, 5.5/4.6/6.5 mm, and 7.4/5.0/6.6 mm at 1, 10, and 20 cm from the center of the FOV, respectively. Measured sensitivity at the center of the FOV (20.84 cps/kBq) was significantly higher than systems with reduced axial FOV. The peak noise-equivalent counting rate was 266.3 kcps at 20.8 kBq/ml, with a corresponding scatter fraction of 40.2%. The correction accuracy for count losses up to the peak noise-equivalent counting rate was 3.6%. For the 10-, 13-, 17-, 22-, 28-, and 37mm spheres, contrast recoveries in the image quality phantom were measured to be 46.2%, 54.3%, 66.1%, 71.1%, 85.3%, and 89.3%, respectively. The mean energy and timing resolution were 9.55% and 381.7 ps, respectively. Phantom and patient images demonstrated excellent image quality, even at short acquisition times or low injected activity. Conclusion: Compared to other PET/CT models, the extended axial FOV improved the overall PET performance of the 5-ring GE Discovery MI scanner. This system offers the potential to reduce scan times or injected activities through increased sensitivity.
Oxygen homeostasis is important in the regulation of biological function. Disease progression can be monitored by measuring oxygen levels, thus producing information for the design of therapeutic treatments. Non-invasive measurements of tissue oxygenation require the development of tools with minimal adverse effects and facile detection of features of interest. Fluorine magnetic resonance imaging (19F-MRI) exploits the intrinsic properties of perfluorocarbon (PFC) liquids for anatomical imaging, cell tracking, and oxygen sensing. However, the highly hydrophobic and lipophobic properties of perfluorocarbons require the formation of emulsions for biological studies. Though, stabilizing these emulsions has been challenging. To enhance the stability and biological loading of perfluorocarbons, one option is to incorporate perfluorocarbon liquids into the internal space of biocompatible mesoporous silica nanoparticles. Here, we developed perfluorocarbon-loaded ultraporous mesostructured silica nanoparticles (PERFUMNs) as 19F-MRI detectable oxygen sensing probes. Ultraporous mesostructured nanoparticles (UMNs) have large internal cavities (average = 1.76 cm3 g−1), facilitating an average 17% loading efficiency of PFCs, meeting the threshold fluorine concentrations needed for imaging studies. Perfluoro-15-crown-5-ether PERFUMNs have the highest equivalent nuclei per PFC molecule, and a spin-lattice (T1) relaxation-based oxygen sensitivity of 0.0032 mmHg−1 s−1 at 16.4 T (657 MHz). The option of loading PFCs after synthesizing UMNs, rather than the more traditional in situ core-shell syntheses, allows for use of a broad range of PFC liquids from a single material. The biocompatible and tunable chemistry of UMNs combined with the intrinsic properties of PFCs makes PERFUMNs a MRI sensor with potential for anatomical imaging, cell tracking, and metabolic spectroscopy with improved stability.
19F MRI is valuable for in vivo imaging due to the only trace amounts of fluorine in biological systems. Because of the low sensitivity of MRI however, designing new fluorochemicals remains a significant challenge for achieving sufficient 19F signal. Here, we describe a new class of high-signal, water-soluble fluorochemicals as 19F MRI imaging agents. A polyamide backbone is used for tuning the proteolytic stability to avoid retention within the body, which is a limitation of current state-of-the-art perfluorochemicals. We show that unstructured peptides containing alternating N-ε-trifluoroacetyllysine and lysine provide a degenerate 19F NMR signal. 19F MRI phantom images provide sufficient contrast at micromolar concentrations, showing promise for eventual clinical applications. Finally, the degenerate high signal characteristics were retained when conjugated to a large protein, indicating potential for in vivo targeting applications, including molecular imaging and cell tracking.
Our aim is to evaluate in phantom and patient studies a recently developed elastic motion debluring (EMDB) technique which makes use of all the acquired PET data and compare its performance to other conventional techniques such as phase based gating (PBG) and HDChest (HDC) both of which use fractions of the acquired data. Comparisons were made with respect to static whole-body (SWB) images with no motion correction. A phantom simulating respiratory motion of the thorax with lung lesions (5 spheres with ID=10- 28 mm) was scanned with 0, 1, 2, and 3 cm motion. Four reconstructions were performed: SWB, PBG, HDC, and EMDB. For PBG, the average (PBGave) and maximum bin (PBGmax) were used. To compare the reconstructions, the ratios of SUV (RSmax), SUVpeak (RSpeak), and CNR (RCNR) were calculated with respect to SWB. Additionally, 46 patients with lung or liver tumors < 3 cm diameter were also studied. Measurements of SUV, SUVpeak, and contrast-to-noise ratio (CNR) were made for 46 lung and 19 liver lesions. To evaluate image noise, the SUV standard deviation was measured in healthy lung and liver tissue and in the phantom background. Finally, subjective image quality of patient exams was scored on a 5 point scale by four radiologists. In the phantom, EMDB increased SUV/SUVpeak over SWB but to a lesser extent than the other reconstruction methodologies. The RCNR for EMDB however was higher than all other reconstructions (0.68 EMDB > 0.54 HDC > 0.41 PBGmax > 0.31 PBGave). Similar results were seen in patient studies. The SUV/SUVpeak were higher by 19.3/11.1% EMDB, 21.6/13.9% HDC, 22.8/12.8% PBGave, and 45.6/26.8% PBGmax compared to SWB. Lung/liver noise increased EMDB (3/15%), HDC (35/56%), PBGave (100/170%), and PBGmax (146/219%). CNR increased in lung/liver tumors only for EMDB (18/13%), and decreased for HDC (-14/-23%), PBGave (-39/-63%), and PBGmax (-18/-46%). Average radiologist scores of image quality were SWB (4.0 ± 0.8) > EMDB (3.7 ± 1.0) > HDC (3.1 ± 1.0) > PBG (1.5 ± 0.7). The EMDB algorithm had the least increase in image noise, improved lesion CNR, and had the highest overall image quality score.
Artifacts in digital breast tomosynthesis and synthesized 2D imaging can obscure important findings on mammograms. The radiologist, mammography technologist, and medical physicist must be able to recognize these artifacts and use the vendor's new processing algorithms to mitigate the effects of such artifacts.
Techniques to monitor the oxygen partial pressure (pO 2 ) within implanted tissue-engineered grafts (TEGs) are critically necessary for TEG development, but current methods are invasive and inaccurate. In this study, we developed an accurate and noninvasive technique to monitor TEG pO 2 utilizing proton ( 1 H) or fluorine ( 19 F) magnetic resonance spectroscopy (MRS) relaxometry. The value of the spin-lattice relaxation rate constant (R 1 ) of some biocompatible compounds is sensitive to dissolved oxygen (and temperature), while insensitive to other external factors. Through this physical mechanism, MRS can measure the pO 2 of implanted TEGs. We evaluated six potential MRS pO 2 probes and measured their oxygen and temperature sensitivities and their intrinsic R 1 values at 16.4 T. Acellular TEGs were constructed by emulsifying porcine plasma with perfluoro-15-crown-5-ether, injecting the emulsion into a macroencapsulation device, and cross-linking the plasma with a thrombin solution. A multiparametric calibration equation containing R 1 , pO 2 , and temperature was empirically generated from MRS data and validated with fiber optic (FO) probes in vitro. TEGs were then implanted in a dorsal subcutaneous pocket in a murine model and evaluated with MRS up to 29 days postimplantation. R 1 measurements from the TEGs were converted to pO 2 values using the established calibration equation and these in vivo pO 2 measurements were simultaneously validated with FO probes. Additionally, MRS was used to detect increased pO 2 within implanted TEGs that received supplemental oxygen delivery. Finally, based on a comparison of our MRS data with previously reported data, ultra-high-field (16.4 T) is shown to have an advantage for measuring hypoxia with 19 F MRS. Results from this study show MRS relaxometry to be a precise, accurate, and noninvasive technique to monitor TEG pO 2 in vitro and in vivo.
Measurements of standardized uptake values (SUV) can vary due to many causes, including respiratory motion. Various methodologies have been introduced to correct for motion in PET, with quiescent-period-gated (QPG) PET being the most popular approach. QPG has been shown to improve PET image quantification compared to static-whole-body (SWB) PET. However, to achieve this improvement, QPG PET requires CT attenuation correction data that matches the QPG PET data. In this paper we investigated the effect of using free-breathing CT for attenuation correction of QPG PET on SUVmax and SUVpeak and compared the results to those of SWB PET. 34 lesions in 27 patients were included. All patients were injected with F-18 FDG. 4D-CT datasets representing all possible phases of respiration that could result from a free-breathing CT were acquired. The 4D-CT datasets were used for attenuation correction of the QPG and SWB PET data. Percentage change in the SUVmax and SUVpeak range was calculated for the reconstructions and compared between QPG and SWB PET. The mean percentage change in the lesion SUVmax and SUVpeak ranges were 19.1% (p = 0.0178) and 25.2% (p = 0.0002) higher for QPG compared to SWB, respectively. The maximum percent change in SUVmax and SUVpeak ranges were 58.5% and 59.0% for QPG, respectively compared to 46.1% and 45.3% for SWB, respectively. The highest SUVmax and SUVpeak measurements corresponded to the CT phase that matched the QPG phase. Utilizing free-breathing CT for attenuation correction can lead to large changes in quantification due to misalignment with PET data. This misalignment has a large quantitative impact on QPG PET as compared to SWB PET. When interpreting quantitative changes in lesions, it is critical to consider the influences of free-breathing CT-based attenuation correction.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.