Abstract:In this research, a finite element study on a nanoneedle-microfluidic system for single cell temperature measurement is presented. The nanoneedle design and electrical and mechanical characterization are analyzed, in which tungsten is used as the sensing material. A rectangular shaped sensor with a gap of 10.8 µm showed to give the same current density distribution within the nanoneedle, and a 90 nm 2 cross-sectional area showed to cause minimum damage to the cell. Furthermore, the current showed to have a positive temperature coefficient of resistance (TCR) with an increase in the temperature, and the nanoneedle showed to be able to resist ramp force up to 22.5 µN before failure. Electrical measurement on yeast cell showed that the nanoneedle was independent of the cell conductivity. The nanoneedle proved to be able to measure temperature with a current difference of 50 nA and a resolution of 0.02 • C in 10 ms. A Y-shaped microchannel was proposed and the microchannel cross-sectional area was optimized to be 63 µm 2 and a flow rate of 24.6 pL/min allowed successful cell penetration causing minimal damage to the cell.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.