The tensioners tested achieved significantly different wire tension values than those reported by the manufacturer. This discrepancy could lead to under-tensioning and allowing excessive movement at a fracture site or over-tensioning, leading to wire breakage. We recommend tensioning wires at least to the recommended line on the device for 66 mm and 84 mm rings and at most to the recommended line for 118 mm rings. Further studies are needed to evaluate other veterinary wire tensioners and to develop a calibration method for these devices in practice.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.