Purpose: To compare the axial length (AL) obtained by A-scan biometry (PAC SCAN 300AP; Sonomed Escalon, USA) and LENSTAR-LS 900 (Haag-Streit, Koeniz, Switzerland) in silicone oil (SiO)-filled eyes. Methods: AL measurements were taken in 50 SiO-filled eyes using A-scan and LENSTAR-LS 900 before SiO removal and 1 month following SiO removal. In the subset of patients requiring intraocular lens (IOL) insertion, the predicted refraction and the refraction obtained were compared. IOL power in these patients was calculated using SRK-T formula and the AL obtained by LENSTAR. Results: In SiO-filled eyes, a significant difference was noted between the AL values obtained using the two methods (P = 0.0002). No significant difference was noted after SiO removal (P = 0.634). In the subset of patients needing IOL insertion, no significant difference (P = 0.07) was seen between target refractive error and postoperative refractive error (mean deviation from the target being 0.176 diopter). AL of an SiO-filled eye is more accurately measured using optical low coherence reflectometry (OLCR)-based biometry (LENSTAR) than with conventional acoustic biometry (A-scan). Conclusion: We conclude that LENSTAR gives more accurate biometry in an SiO-filled eye. The AL obtained after SiO removal was comparable and showed no significant difference.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.