AIMTo investigate viability assessment of segmental small bowel ischemia/reperfusion in a porcine model.METHODSIn 15 pigs, five or six 30-cm segments of jejunum were simultaneously made ischemic by clamping the mesenteric arteries and veins for 1 to 16 h. Reperfusion was initiated after different intervals of ischemia (1-8 h) and subsequently monitored for 5-15 h. The intestinal segments were regularly photographed and assessed visually and by palpation. Intraluminal lactate and glycerol concentrations were measured by microdialysis, and samples were collected for light microscopy and transmission electron microscopy. The histological changes were described and graded.RESULTSUsing light microscopy, the jejunum was considered as viable until 6 h of ischemia, while with transmission electron microscopy the ischemic muscularis propria was considered viable until 5 h of ischemia. However, following ≥ 1 h of reperfusion, only segments that had been ischemic for ≤ 3 h appeared viable, suggesting a possible upper limit for viability in the porcine mesenteric occlusion model. Although intraluminal microdialysis allowed us to closely monitor the onset and duration of ischemia and the onset of reperfusion, we were unable to find sufficient level of association between tissue viability and metabolic markers to conclude that microdialysis is clinically relevant for viability assessment. Evaluation of color and motility appears to be poor indicators of intestinal viability.CONCLUSIONThree hours of total ischemia of the small bowel followed by reperfusion appears to be the upper limit for viability in this porcine mesenteric ischemia model.
The standard clinical method for the assessment of viability in ischemic small intestine is still visual inspection and palpation. This method is non-specific and unreliable, and requires a high level of clinical experience. Consequently, viable tissue might be removed, or irreversibly damaged tissue might be left in the body, which may both slow down patient recovery. Impedance spectroscopy has been used to measure changes in electrical parameters during ischemia in various tissues. The physical changes in the tissue at the cellular and structural levels after the onset of ischemia lead to time-variant changes in the electrical properties. We aimed to investigate the use of bioimpedance measurement to assess if the tissue is ischemic, and to assess the ischemic time duration. Measurements were performed on pigs (n = 7) using a novel two-electrode setup, with a Solartron 1260/1294 impedance gain-phase analyser. After induction of anaesthesia, an ischemic model with warm, full mesenteric arterial and venous occlusion on 30 cm of the jejunum was implemented. Electrodes were placed on the serosal surface of the ischemic jejunum, applying a constant voltage, and measuring the resulting electrical admittance. As a control, measurements were done on a fully perfused part of the jejunum in the same porcine model. The changes in tan δ (dielectric parameter), measured within a 6 h period of warm, full mesenteric occlusion ischemia in seven pigs, correlates with the onset and duration of ischemia. Tan δ measured in the ischemic part of the jejunum differed significantly from the control tissue, allowing us to determine if the tissue was ischemic or not (P < 0.0001, F = (1,75.13) 188.19). We also found that we could use tan δ to predict ischemic duration. This opens up the possibility of real-time monitoring and assessment of the presence and duration of small intestinal ischemia.
Measurements were performed using a two-electrode setup, with a Solartron 1260/1294 impedance gain-phase analyser. Electrodes were placed on the surface of ischemic pig jejunum, applying a voltage and measuring the resulting electrical admittance. In each pig, 4 segments of the jejunum were made ischemic by clamping the mesenteric arteries and veins, resulting in a 30 cm central zone of warm ischemia and edema. The in vivo part of the experiment lasted 10 h, after which 3 pieces of perfused small intestine were resected, stored in Ringer-acetat at 38 °C, and measured during a 10 h ex vivo experiment. Main results and significance: We found significant differences (p < 0.0001) between the values of electric parameters when comparing the in vivo and ex vivo measurements as a function of ischemic time development. We also observed some similarities in the trends. In vivo, we measured an overall decrease in impedance during the duration of the experiment, probably as a result from the formation of edema. Ex vivo, the low frequency impedance increased initially for approximately 3 h before starting to decrease.
BackgroundOxygen delivering modalities like humidified high-flow nasal cannula (HFNC) and non-invasive positive-pressure ventilation (NIV) are suspected of generating aerosols that may contribute to transmission of disease such as COVID-19. We sought to assess if these modalities lead to increased aerosol dispersal compared to the use of non-humidified low-flow nasal cannula oxygen treatment (LFNC).MethodsAerosol dispersal from 20 healthy volunteers using HFNC, LFNC and NIV oxygen treatment was measured in a controlled chamber. We investigated effects related to coughing and using a surgical facemask in combination with the oxygen delivering modalities. An aerodynamic particle sizer measured aerosol particles (APS3321, 0.3–20 µm) directly in front of the subjects, while a mesh of smaller particle sensors (SPS30, 0.3–10 µm) was distributed in the test chamber.ResultsNon-productive coughing led to significant increases in particle dispersal close to the face when using LFNC and HFNC but not when using NIV. HFNC or NIV did not lead to a statistically significant increase in aerosol dispersal compared to LFNC. With non-productive cough in a room without air changes, there was a significant drop in particle levels between 100 cm and 180 cm from the subjects.ConclusionsOur results indicate that using HFNC and NIV does not lead to increased aerosol dispersal compared to low-flow oxygen treatment, except in rare cases. For a subject with non-productive cough, NIV with double-limb circuit and non-vented mask may be a favourable choice to reduce the risk for aerosol spread.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.