In the past decades, image-guided surgery has evolved rapidly. In procedures with a relatively fixed target area, like neurosurgery and orthopedics, this has led to improved patient outcomes. In cancer surgery, intraoperative guidance could be of great benefit to secure radical resection margins since residual disease is associated with local recurrence and poor survival. However, most tumor lesions are mobile with a constantly changing position. Here, we present an innovative technique for real-time tumor tracking in cancer surgery. In this study, we evaluated the feasibility of real-time tumor tracking during rectal cancer surgery. The application of real-time tumor tracking using an intraoperative navigation system is feasible and safe with a high median target registration accuracy of 3 mm. This technique allows oncological surgeons to obtain real-time accurate information on tumor location, as well as critical anatomical information. This study demonstrates that real-time tumor tracking is feasible and could potentially decrease positive resection margins and improve patient outcome.
The presented surgical navigation system improved the intra-operative awareness about tumor position and orientation, with the potential to improve surgical outcomes for non-palpable breast tumors. Results are positive, and participating surgeons were enthusiastic, but extended surgical experience on real breast tissue is required.
This study shows that transponder tracking is promising for accurate tumor tracking in the operating room. This applies when using the standard FoV, but also when using the extended FoV up to 25 cm above the TA, substantially increasing flexibility.
Purpose Surgical navigation systems are generally only applied for targets in rigid areas. For non-rigid areas, real-time tumor tracking can be included to compensate for anatomical changes. The only clinically cleared system using a wireless electromagnetic tracking technique is the Calypso ® System (Varian Medical Systems Inc., USA), designed for radiotherapy. It is limited to tracking maximally three wireless 5-degrees-of-freedom (DOF) transponders, all used for tumor tracking. For surgical navigation, a surgical tool has to be tracked as well. In this study, we evaluated whether accurate 6DOF tumor tracking is possible using only two 5DOF transponders, leaving one transponder to track a tool. Methods Two methods were defined to derive 6DOF information out of two 5DOF transponders. The first method uses the vector information of both transponders (TTV), and the second method combines the vector information of one transponder with the distance vector between the transponders (OTV). The accuracy of tracking a rotating object was assessed for each method mimicking clinically relevant and worst-case configurations. Accuracy was compared to using all three transponders to derive 6DOF (Default method). An optical tracking system was used as a reference for accuracy. Results The TTV method performed best and was as accurate as the Default method for almost all transponder configurations (median errors < 0.5°, 95% confidence interval < 3°). Only when the angle between the transponders was less than 2°, the TTV method was inaccurate and the OTV method may be preferred. The accuracy of both methods was independent of the angle of rotation, and only the OTV method was sensitive to the plane of rotation. Conclusion These results indicate that accurate 6DOF tumor tracking is possible using only two 5DOF transponders. This encourages further development of a wireless EM surgical navigation approach using a readily available clinical system.
Background
Surgical navigation systems generally require intraoperative steps, such as intraoperative imaging and registration, to link the system to the patient anatomy. Because this hampers surgical workflow, we developed a plug-and-play wireless navigation system that does not require any intraoperative steps. In this ex vivo study on human hepatectomy specimens, the feasibility was assessed of using this navigation system to accurately resect a planned volume with small margins to the lesion.
Methods
For ten hepatectomy specimens, a planning CT was acquired in which a virtual spherical lesion with 5 mm margin was delineated, inside the healthy parenchyma. Using two implanted trackers, the real-time position of this planned resection volume was visualized on a screen, relative to the used tracked pointer. Experienced liver surgeons were asked to accurately resect the nonpalpable planned volume, fully relying on the navigation screen. Resected and planned volumes were compared using CT.
Results
The surgeons resected the planned volume while cutting along its border with a mean accuracy of − 0.1 ± 2.4 mm and resected 98 ± 12% of the planned volume. Nine out of ten resections were radical and one case showed a cut of 0.8 mm into the lesion. The sessions took approximately 10 min each, and no considerable technical issues were encountered.
Conclusions
This ex vivo liver study showed that it is feasible to accurately resect virtual hepatic lesions with small planned margins using our novel navigation system, which is promising for clinical applications where nonpalpable hepatic metastases have to be resected with small resection margins.
Members prof.dr. I.A.M.J. Broeders University of Twente prof.dr.ir. C.H. Slump University of Twente dr.ir. F. van der Heijden University of Twente prof.dr. G.L. Beets University of Maastricht prof.dr. H.J.T. Rutten University of Maastricht dr.ir. H.C. Groen Antoni van Leeuwenhoek -Netherlands Cancer InstituteTABLE OF CONTENS
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.