SummaryArterial cannulation is associated with complications including bacterial contamination, accidental intra-arterial injection and blood spillage. We performed a series of audits and experiments to gauge the potential for these, as well as assess the possible contribution of a new device, the Needle-Free Arterial Non-Injectable Connector (NIC), in reducing these risks. The NIC comprises a needle-free connector that prevents blood spillage and a one-way valve allowing aspiration only; once screwed onto the side port of a three-way tap, the device can only be removed with difficulty. We performed a clinical audit of arterial monitoring systems in our intensive care unit, which showed an incidence of bacterial colonisation of five in 86 (6%) three-way tap ports. We constructed a manikin simulation experiment of the management of acute bradycardia, in which trainee doctors were required to inject atropine intravenously. Ten of 15 (66%) doctors injected the drug into the three-way tap of the arterial monitoring system rather than into the intravenous cannula or the central venous catheter. In a laboratory study, we replicated the arterial blood sampling and flushing sequence from a three-way tap, with the syringes attached either directly to the threeway tap port or to a NIC attached to the port. The first (discard) syringe attached to the three-way tap was contaminated with bacteria. Bacterial growth was found in 17 of 20 (85%) downstream flushed samples (corresponding to the patient's circulation) when the three-way tap was accessed directly, compared to none of 20 accessed via the NIC (p < 0.0001). Growth was found on all of 20 (100%) ports accessed directly compared to none of 20 accessed via the NIC (p < 0.0001). The NIC effectively prevents bacteria from contaminating sampling lines. As its design also prevents accidental intra-arterial injection, we suggest that it can reduce complications of arterial monitoring.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.