This guidance document was prepared on behalf of the International Council for Standardization in Haematology (ICSH) for providing haemostasis-related guidance documents for clinical laboratories. This inaugural coagulation ICSH document was developed by an ad hoc committee, comprised of international clinical and laboratory direct acting oral anticoagulant (DOAC) experts. The committee developed consensus recommendations for laboratory measurement of DOACs (dabigatran, rivaroxaban, apixaban and edoxaban), which would be germane for laboratories assessing DOAC anticoagulation. This guidance document addresses all phases of laboratory DOAC measurements, including pre-analytical (e.g. preferred time sample collection, preferred sample type, sample stability), analytical (gold standard method, screening and quantifying methods) and post analytical (e.g. reporting units, quality assurance). The committee addressed the use and limitations of screening tests such as prothrombin time, activated partial thromboplastin time as well as viscoelastic measurements of clotting blood and point of care methods. Additionally, the committee provided recommendations for the proper validation or verification of performance of laboratory assays prior to implementation for clinical use, and external quality assurance to provide continuous assessment of testing and reporting method.
The PFA-100 system is a platelet function analyzer designed to measure platelet-related primary hemostasis. The instrument uses two disposable cartridges: a collagen/epinephrine (CEPI) and a collagen/ADP (CADP) cartridge. Previous experience has shown that CEPI cartridges detect qualitative platelet defects, including acetylsalicylic acid (ASA)-induced abnormalities, while CADP cartridges detect only thrombocytopathies and not ASA use. In this seven-center trial, 206 healthy subjects and 176 persons with various platelet-related defects, including 127 ASA users, were studied. The platelet function status was determined by a platelet function test panel. Comparisons were made as to how well the defects were identified by the PFA-100 system and by platelet aggregometry. The reference intervals for both cartridges, testing the 206 healthy subjects, were similar to values described in smaller studies in the literature [mean closure time (CT) 132 s for CEPI and 93 s for CADP]. The use of different lot numbers of cartridges or duplicate versus singleton testing revealed no differences. Compared with the platelet function status, the PFA-100 system had a clinical sensitivity of 94.9% and a specificity of 88.8%. For aggregometry, a sensitivity of 94.3% and a specificity of 88.3% were obtained. These values are based on all 382 specimens. A separate analysis of sensitivity by type of platelet defect, ASA use versus congenital thrombocytopathies, revealed for the PFA-100 system a 94.5% sensitivity in identifying ASA users and a 95.9% sensitivity in identifying the other defects. For aggregometry, the values were 100% for ASA users and 79.6% for congenital defects. Analysis of concordance between the PFA-100 system and aggregometry revealed no difference in clinical sensitivity and specificity between the systems (p > 0.9999). The overall agreement was 87.5%, with a Kappa index of 0.751. The two tests are thus equivalent in their ability to identify normal and abnormal platelet defects. Testing 126 subjects who took 325 mg ASA revealed that the PFA-100 system (CEPI) was able to detect 71.7% of ASA-induced defects with a positive predictive value of 97.8%. The overall clinical accuracy of the system, calculated from the area under the ROC curve, was 0.977. The data suggest that the PFA-100 system is highly accurate in discriminating normal from abnormal platelet function. The ease of operation of the instrument makes it a useful tool to use in screening patients for platelet-related hemostasis defects.
S. Performance of coagulation tests in patients on therapeutic doses of dabigatran: a cross-sectional pharmacodynamic study based on peak and trough plasma levels. J Thromb Haemost 2013; 11: 1493-1502.Summary. Background: Knowledge of anticoagulation status during dabigatran therapy may be desirable in certain clinical situations. Objective: To determine the coagulation tests that are most useful for assessing dabigatran's anticoagulant effect. Methods: Peak and trough blood samples from 35 patients taking dabigatran 150 mg twice daily, and one sample each from 30 non-anticoagulated individuals, were collected. Mass spectrometry and various coagulation assays were performed. 'Therapeutic range' was defined as the range of plasma dabigatran concentrations determined by mass spectrometry between the 2.5th and 97.5th percentiles of all values. Results: The therapeutic range was 27-411 ng mL
À1. The prothrombin time (PT) and activated partial thromboplastin time (APTT), determined with multiple reagents, and activated clotting time (ACT) were insensitive to therapeutic dabigatran: 29%, 18% and 40% of samples had a normal PT, APTT, and ACT, respectively. However, normal PT, ACT and APTT ruled out dabigatran levels above the 75th percentile. The thrombin clotting time (TCT) correlated well and linearly with dabigatran levels below the 50th percentile, but was unmeasurable above it. The dilute thrombin time, ecarin clotting time and ecarin chromogenic assay showed linear correlations with dabigatran levels over a broad range, and identified therapeutic and supratherapeutic levels. Conclusions: The prothrombin time, APTT and ACT are often normal in spite of therapeutic dabigatran plasma levels. The TCT is useful for detecting minimal dabigatran levels. The dilute thrombin time and chromogenic and clotting ecarin assays accurately identify therapeutic and supratherapeutic dabigatran levels. This trial is registered at www.clinicaltrials.gov (#NCT01588327).
Cocoa consumption suppressed ADP- or epinephrine-stimulated platelet activation and platelet microparticle formation. Cocoa consumption had an aspirin-like effect on primary hemostasis.
Severe injury usually results in increased platelet activation and function. However, the combination of increased platelet activation with decreased function was associated with increased mortality.
Direct oral anticoagulant therapies, including direct anti-Xa and thrombin inhibitors have recently been introduced and may have advantages over vitamin K antagonists such as warfarin. This review describes briefly the clinical utility and mechanism of action of these agents. Detailed information is provided on effect of these agents on routine assays including the APTT and PT as well as their impact on specialty laboratory assays. Also included are the use of drug specific assays and a discussion of alternative methods to determine relative drug concentration, such as evaluating drug calibrators in APTT and PT assays and using heparin calibrated anti-Xa assays to measure direct Xa inhibitors.
In 2018, the International Council for Standardization in Hematology (ICSH) published a consensus document providing guidance for laboratories on measuring direct oral anticoagulants (DOACs). Since that publication, several significant changes related to DOACs have occurred, including the approval of a new DOAC by the Food and Drug Administration, betrixaban, and a specific DOAC reversal agent intended for use when the reversal of anticoagulation with apixaban or rivaroxaban is needed due to life-threatening or uncontrolled bleeding, andexanet alpha. In addition, this ICSH Working Party recognized areas where additional information was warranted, including patient population considerations and updates in point-of-care testing. The information in this manuscript supplements our previous ICSH DOAC laboratory guidance document. The recommendations provided are based on (1) information from peer-reviewed publications about laboratory measurement of DOACs, (2) contributing author’s personal experience/expert opinion and (3) good laboratory practice.
No PT or APTT reagent system effectively detected apixaban. All anti-Xa methods demonstrated sensitivity to low concentrations of DOAC. Dilute viper venom methods are exquisitely sensitive to anti-Xa DOAC, suggesting potential use of this assay for screening or measuring these drugs.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.