Within the EC-funded CHANCE project several non-destructive techniques are being considered for the assay of waste bearing drums. Such techniques include calorimetry, gamma-ray spectrometry and neutron coincidence counting. The aim is to quantify uncertainties on the inventory of radionuclides, and how these are potentially reduced by combining the signatures from different techniques in the data analysis.
In this framework, neutron coincidence measurements were carried out with two slab counters based on 3He detectors coupled to shift register electronics. Such a system consists of two identical slabs with 6 detectors each, and is transportable, rather compact and flexible in terms of sizes and geometries that can be measured. With this system three 200 L drums containing certified reference nuclear material and different filling materials were measured. The certified nuclear material was in the form of 21 pellets of mixed oxide of U and Pu with a total mass of about 10.5 g; in addition, a single pellet of about 10.05 g was also available. The pellets could be placed in predefined positions within the drum in a reproducible way. The geometry and composition of the three drums was well characterized and consisted of Ethafoam, a mixture of Ethafoam, stainless steel and PVC, and mortar with an inner core of extruded polystyrene. The measurement setup was arranged such that the drum was placed between the two slab counters. The positions of the slab counters relative to the drum were accurately measured before each measurement, and a dedicated system was used to minimize the uncertainty on the detector positioning.
The measurement data were first analysed by applying the point model of Hage and the mass of nuclear material in the drum was determined from the rate of totals and reals and the radionuclide composition. Due to the fact that not all the point model conditions were met, we found that the point model overestimates the mass up to about 50%. In addition, a Monte Carlo model of the measurement geometry was developed using the MCNP code. The model was used to determine a calibration factor between the reals rate and the mass of the sample. Measurements with a calibrated 252Cf source were used to verify the model. With a Monte Carlo based approach the mass of the mixed oxide pellets is within a few percent from the nominal values, except for strongly asymmetrical configurations where the deviation is up to about 20%. The results reveal the importance of an accurate background correction and of accounting for surrounding materials of the building such as walls, floor and ceiling in the Monte Carlo model.
Several non-destructive assays techniques have been developed for the measurement of fissile materials in the fields of dismantling, decommissioning, nuclear security, and nuclear safeguards. Among these techniques, neutron coincidence counting is based on the detection of time-correlated neutrons from induced and spontaneous fissions. 3He Tubes have been the primary choice for neutron coincidence counting due to their high detection efficiency, rather low sensitivity to gamma-rays and proven field reliability. This paper covers the implementation of a new electronic setup to a Canberra WM3400 neutron coincidence counter.
First we describe the properties of the used detectors, with focus on the characteristics of the default electronics and highlight its limitations such as the high input capacitance, short shaping time and the necessity for selected tubes. We then propose the new electronic setup to overcome these limitations. This setup includes a dedicated preamp for every tube , the possibility to adjust for gain differences between the tubes and a better optimised shaping time for 3He detectors.
We carried out measurements with the two electronic systems to compare their performances in terms of gamma-ray sensitivity, efficiency and die-away time. The gamma ray sensitivity was measured with calibrated 137Cs and a 60Co sources at the Laboratory for Nuclear Calibration of the Belgian Nuclear Research Centre with dose rates between 10 μSv/h and 50 mSv/h. Measurements with a 252Cf source were used to determine the die-away time of the system and the total measurement efficiency for the considered geometry.
The measurements showed that, with the default electronics, neutron count-rates are already affected by gamma radiation at a dose rate of 10÷30 μSv/h. On the other hand the neutron coincidence counter equipped with the new electronics proved to be insensitive to gamma-radiation up to a dose rate of at least 20 mSv/h. The high-voltage set with the new electronics is lower than in the case of the default electronics and is within the range recommended by the tubes manufacturer. The die-away time was not affected by the used electronics. A reduction of about 20% in the neutron detection efficiency due to the used discriminator threshold was observed.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.