We treated 34 patients with recurrent dislocation of the hip with a constrained acetabular component. Roentgen stereophotogrammetric analysis was performed to assess migration of the prosthesis. The mean clinical follow-up was 3.0 years (2.2 to 4.8) and the radiological follow-up was 2.7 years (2.0 to 4.8). At the latest review six patients had died and none was lost to follow-up. There were four acetabular revisions, three for aseptic loosening and one for deep infection. Another acetabular component was radiologically loose with progressive radiolucent lines in all Gruen zones and was awaiting revision. The overall rate of aseptic loosening was 11.8% (4 of 34). Roentgen stereophotogrammetric analysis in the non-revised components confirmed migration of up to 1.06 mm of translation and 2.32 degrees of rotation at 24 months. There was one case of dislocation and dissociation of the component in the same patient. Of the 34 patients, 33 (97.1%) had no further episodes of dislocation. The constrained acetabular component reported in our study was effective in all but one patient with instability of the hip, but the rate of aseptic loosening was higher than has been reported previously and requires further investigation.
BackgroundHigh-flow nasal oxygen (HFNO) is frequently used in hospitals, producing droplets and aerosols that could transmit SARS-CoV-2.AimTo determine if a headbox could reduce droplet and aerosol transmission from patients requiring HFNO.MethodsThe size and dispersion of propylene glycol (model for patient-derived infectious particles) was measured using a spectrometer and an infant mannequin receiving 10–50 L/min of HFNO using (1) no headbox, (2) open headbox, (3) headbox-blanket or (4) headbox with a high-efficiency particulate (HEP) filter covering the neck opening.ResultsAll headbox set-ups reduced the dispersal of droplets and aerosols compared with no headbox. The headbox-blanket system increased aerosol dispersal compared with the open headbox. The fraction of aerosols retained in the headbox for HFNO of 10 and 50 L/min was, respectively, as follows: (1) open headbox: 82.4% and 42.2%; (2) headbox-blanket: 56.8% and 39.5%; (3) headbox-HEP filter: 99.9% and 99.9%.ConclusionA HEP-filter modified headbox may serve as an effective droplet and aerosol barrier adjunct for the protection of staff caring for children receiving HFNO.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.