Selective and novel method has been optimized for evaluation of Ethamsylate and Tranexamic acid in bulk, formulation and equipment cleaning samples by UPLC-PDA. The method was developed with the Buffer (pH: 7.0): Acetonitrile, (75:25%, v/v) using the Acquity BEH C 18 ,1.7µ.m,2.1mmX50mm analytical column with analysis time of five minutes. The flow of mobile phase through column was 0.8ml/min.The sample volume was 5 µL. The detection was carried at 260nm in a photodiode array detector. The retention times of Ethamsylate and Tranexamic acid were 1.121 min and 2.122 min respectively. The correlation coefficient (r 2) was 0.9997 with linear range of 2.5-30.0 µg/ml for EtamsylateFor Tranexamic acid 5µg/ml-60.0 µg/ml .The correlation coefficient (r 2) for Tranexamic acid was found to be 1.000. The LOQ for the Ethamsylate and Tranexamic acid was 25ng/ml and 50ng/ml respectively. The developed method was applied for the bulk,formulation and equipment cleaning samples.The results were found satisfactory.Hence the developed method was proved more accurate and sensitive.It can be used at all the stages in pharmaceutical manufacturing and quality control for evaluating of Ethamsylate and Tranexamic acid.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.