Background
Multiple infection outbreaks have been linked to contaminated duodenoscopes worldwide. However, the contamination rate of patient-ready duodenoscopes varies highly amongst published studies testing this subject. We aimed to estimate the contamination rate of reprocessed patient-ready duodenoscopes for endoscopic retrograde cholangio-pancreatography (ERCP) based on currently available data.
Methods
We searched the PubMed and Embase databases from January 1, 2010 until March 10, 2020, for citations investigating contamination rates of reprocessed patient-ready duodenoscopes. Studies not assessing other types of endoscopes than duodenoscopes were excluded from the analysis. Study eligibility and data extraction was evaluated by three reviewers independently. A random-effects model (REM) based on the proportion distribution was used to calculate the pooled total contamination rate of reprocessed patient-ready duodenoscopes. Subgroup analyses were carried out to assess contamination rates when using different reprocessing methods by comparing single high-level disinfection (HLD) with double HLD and ethylene oxide (EtO) gas sterilization. Additionally, we investigated the contamination rate between studies conducted following an outbreak compared to non-outbreak-initiated studies.
Findings
We identified 15 studies that fulfilled the inclusion, which included 925 contaminated duodenoscopes from 13,112 samples. The calculated total weighted contamination rate was 15.25% ± 0.018 (95% confidence interval [Cl]: 11.74% - 18.75%). The contamination rate after only using HLD was 16.14% ± 0.019 (95% Cl: 12.43% - 19.85%) and after using either dHLD or EtO the contamination rate decreased to 9.20% ± 0.025 (95% Cl: 4.30% - 14.10%). Studies conducted following an outbreak (n=4) showed a 5.72% ± 0.034 (95% Cl: 0.00% - 12.43%) contamination rate, and non-outbreak-initiated studies (n=11) revealed a contamination rate of 21.50% ± 0.031 (95% Cl: 15.35% - 27.64%).
Interpretation
This is the first meta-analysis to estimate the contamination rate of patient-ready duodenoscopes used for ERCP. Based on the available literature, our analysis demonstrates that there is a 15.25% contamination rate of reprocessed patient-ready duodenoscopes. Additionally, the analysis indicates that dHLD and EtO reprocessing methods are superior to single HLD but still not efficient in regards to cleaning the duodenoscopes properly. Furthermore, studies conducted following an outbreak did not entail a higher contamination rate compared to non-outbreak-initiated studies.
Funding
The authors received no financial support for the research, authorship, and/or publication of this article.
Background
This study investigated whether the single-use rhinolaryngoscope is clinically and economically comparable to the conventional reusable rhinolaryngoscope within a tertiary otolaryngology centre in the UK.
Methods
A non-blinded, prospective and single-arm evaluation was carried out over a 5-day period, in which micro-costing was used to compare single-use rhinolaryngoscopes with reusable rhinolaryngoscopes.
Results
Overall, 68 per cent of the investigators perceived the single-use rhinolaryngoscope to be ‘good’ or ‘very good’, while 85 per cent believed the single-use rhinolaryngoscope could replace the reusable rhinolaryngoscope (n = 59). The incremental costs of reusable rhinolaryngoscope eyepieces and videoscopes in the out-patient clinic, when compared to single-use rhinolaryngoscopes, were £30 and £11, respectively. The incremental costs of reusable rhinolaryngoscope eyepieces and videoscopes in the acute surgical assessment unit, when compared to single-use rhinolaryngoscopes, were −£4 and −£73, respectively.
Conclusion
The single-use rhinolaryngoscope provides a clinically comparable, and potentially cost-minimising, alternative to the reusable rhinolaryngoscope for use in the acute surgical assessment unit of our hospital.
Background and Objective Until 2009, only reusable bronchoscopes were marketed, but the introduction and widespread adoption of single-use flexible bronchoscopes (SFBs) as an emerging technology has since accelerated. Several studies have described the costs of reusable flexible bronchoscopes (RFBs) and SFBs. This meta-analysis aimed to compile the current published evidence to analyse the cost of different scenarios using RFBs and SFBs. Methods All published literature describing the cost of RFBs or SFBs was identified by searching PubMed, Embase and Google Scholar, limited to those between 1 January, 2009 and 6 November, 2020. Included studies should report the total cost of RFBs. Continuous data were extracted for relevant outcomes and analysed using RStudio ® 4.0.3 as the standardised mean difference and standard error of the mean in a mixed-effects model. Risk of bias was assessed based on the reporting quality. Results In the systematic literature review, 342 studies were initially identified, and 11 were included in the final analysis.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.