SUMMARY
The X‐ray microanalytical spatial resolution is determined experimentally in various analytical electron microscopes by measuring the degradation of an atomically discrete composition profile across an interphase interface in a thin‐foil of Ni‐Cr‐Fe. The experimental spatial resolutions are then compared with calculated values. The calculated spatial resolutions are obtained by the mathematical convolution of the electron probe size with an assumed beam‐broadening distribution and the single‐scattering model of beam broadening. The probe size is measured directly from an image of the probe in a TEM/STEM and indirectly from dark‐field signal changes resulting from scanning the probe across the edge of an MgO crystal in a dedicated STEM. This study demonstrates the applicability of the convolution technique to the calculation of the microanalytical spatial resolution obtained in the analytical electron microscope. It is demonstrated that, contrary to popular opinion, the electron probe size has a major impact on the measured spatial resolution in foils < 150 nm thick.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.