Performing a preuse self-test on a nonexpanded pediatric circuit that is then expanded leads to falsely elevated displayed tidal volume in infants less than 10 kg during pressure-controlled ventilation. Overestimation of reported tidal volume can be avoided by expanding the breathing circuit tubing to the length which will be used during a case prior to performing the anesthesia machine preuse self-test. After department-wide education and implementation, performing a correct preuse self-test is now the standard practice in our cardiac operating rooms.
Children with suspected pulmonary hypertension must undergo extensive and invasive evaluations to establish a definitive diagnosis. A previously healthy 4-year old girl, newly diagnosed with suprasystemic pulmonary hypertension required multiple lung biopsies. Each procedure was associated with significant bleeding. The challenging anesthetic management of lung biopsy in the presence of suprasystemic pulmonary hypertension is described.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.