To evaluate the biocompatibility and the efficacy of leukocyte removal filters, we performed a prospective study by using the cardiopulmonary bypass perfusate taken from the heart鈥恖ung machine for 20 patients who underwent cardiac surgery and were randomly divided into four groups according to the filters used. A leukocyte removal filter was installed in the transfusion line while the perfusate was transfused to the patients. No increases of C3a, C5a, elastase, and thromboxane were found during leukocyte filtration by polyester filters (Optima, Sepacell R500, and Pall RC100). Activation of the complement cascade was observed during filtration by the cellulose acetate filter (Cellselect) although the efficacy of the Cellselect filter was evidently higher than that of the poly鈥恊ster filter. These results imply that polyester leukocyte filters are superior to cellulose acetate filters in terms of biocompatibility but have a reduced efficacy. An optimal leukocyte filter providing both high efficacy and biocompatibility has yet to be developed.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations鈥揷itations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.