To evaluate the accuracy of two non-invasive techniques for cardiac output (CO) measurement, we have measured CO simultaneously by thoracic electrical bioimpedance (TEB), pulsed Doppler ultrasound (DU) and standard thermodilution methods (TD) under different clinical conditions. Measurements were made in 10 patients: (I) during steady state anaesthesia with controlled IPPV ventilation (n = 131), spread over the entire ventilatory cycle; (II) during apnoea (n = 56); (III) during spontaneous breathing (n = 152) in the intensive care unit. Mean (SD) cardiac output values were: (I) COTD 3.5 (1.0) litre min-1, COTEB 3.4 (0.7) litre min-1, CODU 2.8 (0.7) litre min-1; (II) COTD 3.6 (0.6) litre min-1, COTEB 3.5 (0.4) litre min-1, CODU 2.9 (0.7) litre min-1; (III) COTD 7.7 (1.5) litre min-1, COTEB 7.6 (1.9) litre min-1, CODU 5.2 (1.4) litre min-1. The mean percentage deviation of TEB from TD ranged from -2.2% to 1.4% and that of DU from TD was from -16% to -32%. There were no statistically significant differences between TD and TEB, but TD and DU differed significantly during IPPV, apnoea and spontaneous ventilation (P < 0.0001).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.