A validation study was carried out in order to evaluate the efficiency of inductively coupled plasma-optical emission spectrometry (ICP-OES) for the analysis of minor elements (manganese, chromium, copper, iron, and titanium) in aluminium alloys. Aluminium casting samples were obtained by adding compressed powder compacts of each alloying element and aluminium (minitablets) to aluminum baths in a laboratory crucible furnace. Digestion of solid samples was performed using concentrated HCl and H2O2 35% (v/v) previous to analysis by ICP-OES without any matrix separation. This solution-based method was validated considering direct current arc spectrometry as the reference method based on direct analysis without any pretreatment of the solid samples considered. Univariate statistical procedures were carried out, for which precision <3% and trueness of the analytical results were taken into account.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.