The relative pressure difference across stenotic blood vessels serves as an important clinical index for the diagnosis of many cardiovascular diseases. While the clinical gold standard for relative pressure difference measurements is invasive catheterization, Phase-Contrast Magnetic Resonance Imaging has emerged as a promising tool for enabling a noninvasive quantification, by linking highly spatially resolved velocity measurements with relative pressures via the incompressible Navier-Stokes equations. In this work, we provide a review and analysis of current methods for relative pressure estimation and propose 3 additional techniques. Methods are compared using synthetic data from numerical examples, and sensitivity to subsampling and noise was explored. Through our analysis, we verify that the newly proposed approaches are more robust with respect to spatial subsampling and less sensitive to noise and therefore provide improved means for estimating relative pressure differences noninvasively.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.