Objective: This study aimed to develop an analytical method for determining metformin concentrations in human dried blood spots (DBS) using optimal chromatographic conditions and DBS preparation methods and validate the analytical methods.Methods: Analysis metformin HCl performed using C-18 columns (250 mm × 4.6 mm) at a column temperature of 40°C with a mobile phase acetonitrile phosphate buffer (40:60% v/v) at pH 7.0 using a flow rate of 0.8 mL/min, a photodiode array detector at a wavelength of 234 nm, and atorvastatin calcium as an internal standard. Samples were prepared by protein precipitation with 60% methanol and then drying using nitrogen gas at 60°C for 15 min before reconstituting in 200 µL of the mobile phase.Results: Invalidation procedures, the present methods met the acceptance criteria stated in the EMEA Bioanalytical Guideline 2011. The method was also linear at the concentration range of 25.0-5000.0 ng/mL (r=0.9997).
Conclusion:The developed method for determination of metformin HCl in DBS sample was valid with a concentration range of 25.0-5000.0 ng/mL and can be applied to in vivo studies.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.