The PLOS ONE EditorsThe article reports the recording of telemetric data from implanted TL11M2-F20-EET transmitters. These results are described in Figures 4 and 5.The TL11M2-F20-EET transmitter has a reported transmission bandwidth of 1-50Hz and a sampling frequency of 250 Hz. The maximum identifiable signal for the device is 125 Hz.The article reports experiments measuring signals of up to 500Hz. Given the limitations of the transmitter employed, the results reported for frequencies beyond 100Hz are not reliable. We are thus issuing this notification to alert readers that the results in Figure 4 and Figure 5C reporting signals over 100Hz are not reliable.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.