Two methods for the chiral purity determination of bevantolol were developed, namely capillary electrophoresis (CE) using carboxymethyl-beta-cyclodextrin (CM-beta-CD) as a chiral selector and high-performance liquid chromatography (HPLC) using a chiral stationary phase. In the HPLC method, the separation of bevantolol enantiomers was performed on a Chiralpak AD-H column by isocratic elution with n-hexane-ethanol-diethylamine (10:90:0.1, v/v/v) as mobile phase. In the CE method, bevantolol enantiomers were separated on an uncoated fused silica capillary with 50 mM amonium phosphate dibasic adjusted to a pH 6.5 with phosphoric acid containing 15 mM CM-beta-CD as running buffer. Validation data such as linearity, recovery, detection limit, and precision of the two methods are presented. The detection limits of S-(-)-bevantolol were 0.1% and 0.05% for CE and HPLC method, respectively and R-(+)-bevantolol were 0.15% and 0.05% for CE and HPLC method, respectively. There was generally good agreement between the HPLC and CE results.
A method was developed and fully validated for the determination of bevantolol, an adrenergic-receptor blocker, in human plasma. Bevantolol and betaxolol as internal standard (I.S) were extracted from 1 mL of human plasma by solid phase extraction technique using Sep-pak silica cartridge. Chromatographic separation was accomplished under isocratic conditions using a reverse-phase C8 analytical column and mixture of dibasic ammonium phosphate (pH 5.7; 50 mM)-acetonitrile (75:25, v/v) as mobile phase, with a detection wavelength at 220 nm. The method was proved to be specific by testing six different human plasma sources. Linearity was established for the concentration ranges of 40-1600 ng/mL with correlation coefficent of 0.9995. The lower limit of quantification 40 ng/mL with precision of 10.9% as C.V%.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.