This study shows that PPVI is feasible and it improves the haemodynamics in a selected patient collective. Apart from one coronary compression, the rate of complications at short-term follow-up was low. Percutaneous pulmonary valve implantation can be performed by experienced interventionalists with similar results as originally published. The intervention is technically challenging and longer clinical follow-up is needed.
To judge whether an Amplatzer™ Septal Occluder can be used as standard therapy instead of surgery for closure of atrial septal defects we report our experiences in 200 patients. Of these patients, 127 had an atrial septal defect with haemodynamically significant left-to-right shunt, 68 patients a persistent oval foramen after presumed paradoxical embolism, and 5 had a fenestration after Fontan-repair. Mean age was 29.8 years (0.8 to 77.7 years). Body weight ranged from 6.9 to 120.0 kg (mean 51.5 kg). After diagnostic cardiac catheterization, and balloon-sizing of the defect, we implanted Amplatzer™ Septal Occluders with stents of 4 to 28 mm diameter. Follow-up studies were obtained after 48hours, and one, six, and twelve months. Transcatheter closure of the atrial septal defect proved successful in all without any relevant residual shunts. In particular, complete closure was achieved in all patients after presumed paradoxical embolism. The mean period of follow-up is 9–5 months, with a range from 0.4 to 23.5 months, giving a total of 1898 patient months. The occlusion rate after three month was 98.1°. A trivial haemodynamically insignificant residual shunt remained in 1.9° of the patients. Fluoroscopy times ranged from 0 to 43.5 minutes, with a median of 8.7 minutes. The excellent results in the short and medium term in children and adults have resulted in using this device routinely at the present time for closure of central atrial septal defects up to a diameter of 28 mm. Final judgement, however, is only possible after long-term follow-up.
Background-The aims of this study were to validate MRI-derived right ventricular (RV) pressure-volume loops for assessment of RV myocardial contractility and then to apply this technique in patients with chronic RV pressure overload for assessment of myocardial contractility, ventricular pump function, and VA coupling. Methods and Results-Flow-directed catheters were guided under MR fluoroscopy (1.5 T) into the RV for invasive pressure measurements. Simultaneously, ventricular volumes and myocardial mass were assessed from cine MRI. From sampled data, RV pressure-volume loops were constructed, and maximal ventricular elastance indexed to myocardial mass (E max_i ) was derived by use of a single-beat estimation method. This MRI method was first validated in vivo (6 swine), with conductance techniques used as reference. Bland-Altman test showed good agreement between methods (E max_i ϭ5.1Ϯ0.5 versus 5.8Ϯ0.7 mm Hg · mL Ϫ1 · 100 g
Covered CP stents may be used as the therapy of choice in patients with complications after CoA repairs, whereas they provide a safe alternative to conventional stenting in patients with severe and complex CoA lesions or advanced age.
IntroductionAcute kidney injury (AKI) is a frequent complication after cardiac surgery with cardiopulmonary bypass in infants. Renal near-infrared spectroscopy (NIRS) is used to evaluate regional oximetry in a non-invasive continuous real-time fashion, and reflects tissue perfusion. The aim of this study was to evaluate the relationship between renal oximetry and development of AKI in the operative and post-operative setting in infants undergoing cardiopulmonary bypass surgery.MethodsIn this prospective study, we enrolled 59 infants undergoing cardiopulmonary bypass surgery for congenital heart disease for univentricular (n = 26) or biventricular (n = 33) repair. Renal NIRS was continuously measured intraoperatively and for at least 24 hours postoperatively and analysed for the intraoperative and first 12 hours, first 24 hours and first 48 hours postoperatively. The renal oximetry values were correlated with the paediatric risk, injury, failure, loss, end (pRIFLE) classification for AKI, renal biomarkers and the postoperative course.ResultsTwenty-eight (48%) infants developed AKI based on pRIFLE classification. Already during intraoperative renal oximetry and further in the first 12 hours, 24 hours and 48 hours postoperatively, significantly lower renal oximetry values in AKI patients compared with patients with normal renal function were recorded (P < 0.05). Of the 28 patients who developed AKI, 3 (11%) needed renal replacement therapy and 2 (7%) died. In the non-AKI group, no deaths occurred. Infants with decreased renal oximetry values developed significantly higher lactate levels 24 hours after surgery. Cystatin C was a late parameter of AKI, and neutrophil gelatinase-associated lipocalin values were not correlated with AKI occurrence.ConclusionOur results suggest that prolonged low renal oximetry values during cardiac surgery correlate with the development of AKI and may be superior to conventional biochemical markers. Renal NIRS might be a promising non-invasive tool of multimodal monitoring of kidney function and developing AKI in infants undergoing cardiac surgery with cardiopulmonary bypass.
Transcatheter closure of perimembranous ventricular septal defects with coils or devices designed to close other lesions may be complicated by embolization or aortic insufficiency. A new asymmetric Amplatzer perimembranous ventricular septal occluder and delivery system was specifically designed for perimembranous defects. This report describes the first use of this device in 27 patients. Implantation was successful in 25 (93%), with 1 removed for device-related aortic insufficiency and inability to position the delivery sheath in another. Device orientation was excellent when the device was initially advanced through a standard delivery sheath positioned in the left ventricular apex. Twenty-three had complete occlusion within 1 week (92%), with a tiny (< 2 mm) residual shunt in the other two. In the 25 subjects with the device left in place, device-related aortic or tricuspid insufficiency, arrhythmias, and embolization were not observed. These excellent acute results need to be confirmed by long-term follow-up.
Objective: To evaluate the feasibility and usefulness of the Cheatham platinum (CP) stent in a broad spectrum of lesions. Methods: Retrospective analysis of 60 implanted CP stents (11-80 mm lengths, 12 covered) between September 2001 and March 2004. Patients: 53 patients aged 2.5-68 years (median 17 years). Body weight ranged from 12-95 kg (median 52 kg). Thirty six patients had aortic (re)coarctation; seven of them had functionally interrupted aortic arches. Thirteen patients had pulmonary artery stenosis and four had stenosis of caval veins or conduits in a total cavopulmonary connection (TCPC). Results: Arterial pressure gradients dropped from 33 mm Hg (range 20-80 mm Hg) to 5 mm Hg (range 0-10 mm Hg) and pressure gradients in TCPC or caval veins dropped from 4 mm Hg (range 4-20 mm Hg) to 0 mm Hg (range 0-3 mm Hg). All stents were placed in the target lesion without complications. Three stent fractures without clinical instability were noted. Conclusions: The CP stent is suitable for the treatment of vessel stenosis in congenital heart diseases from childhood to adulthood. Whether these good results will be stable in the long term needs to be investigated.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.