Aim: To assess the feasibility of a prototype insulin infusion set (IIS) for extended wear in adults with type 1 diabetes.
Materials and Methods:The prototype Capillary Biomedical investigational extended-wear IIS (CBX IIS) incorporates a soft, flexible, reinforced kink-resistant angled nylon-derivative cannula with one distal and three proximal ports to optimize insulin delivery. Twenty adult participants with type 1 diabetes established on insulin pump therapy used the CBX IIS for two 7-day test periods while wearing a Dexcom G5 continuous glucose monitor.Results: Participants were able to wear the CBX IIS for an average of 6.6 ± 1.4 days.Eighty-eight percent (36 of 41) of sets were worn for 7 days. No serious adverse events were reported. Five infusion sets failed prematurely because of: unresolvable hyperglycaemia (three); hyperglycaemia with elevated ketones (one); or infection (one). Median time in range (3.9-10.0 mmol/L) was 62% (54-76). Average glucose levels per day of infusion set wear showed a statistically significant increase over time (p < .001).Conclusions: Our preliminary observations confirm the tolerability of the prototype CBX IIS for extended wear, albeit with a deterioration in glucose control after the third day.
The GluCath System exhibited acceptable accuracy when deployed in a radial artery for up to 48 hours in ICU patients after elective cardiac surgery. Accuracy of venous deployment was substantially lower with significant rates of intravascular thrombus observed using ultrasound.
The IA-CGM system directly measured arterial blood glucose and did not interfere with clinical care. However, accuracy was similar to that of the less invasive SC-CGM device.
ObjectiveThis study investigated the effects of the inflammatory tissue response (ITR) to an insulin infusion set (IIS) on insulin bolus spread over wear time, as well as the effect of cannula insertion angle on the ITR, bolus shape, and pump tubing pressure.Research design and methodsAngled or straight IISs were inserted every other day for 14 days into the subcutaneous tissue of 11 swine and insulin was delivered continuously. Prior to euthanasia, a 70 µL bolus of insulin/X-ray contrast agent was infused while recording a pressure profile (peak tubing pressure, pmax; area under the pressure curve, AUC), followed by the excision of the tissue-catheter specimen. Bolus surface area (SA) and volume (V) were assessed via micro-CT. Tissue was stained to analyze total area of inflammation (TAI) and inflammatory layer thickness (ILT) surrounding the cannula.ResultsA bolus delivered through an angled IIS had a larger mean SA than a bolus delivered through a straight cannula (314.0±84.2 mm2 vs 229.0±99.7 mm2, p<0.001) and a larger volume (198.7±66.9 mm3 vs 145.0±65.9 mm3, p=0.001). Both decreased significantly over wear time, independent of angle. There was a significant difference in TAI (angled, 9.1±4.0 mm2 vs straight, 14.3±8.6 mm2, p<0.001) and ILT (angled, 0.7±0.4 vs straight, 1.2±0.7 mm, p<0.001). pmax (p=0.005) and AUC (p=0.014) were lower using angled IIS. As ILT increased, pmax increased, while SA and V decreased.ConclusionsThe progression of the ITR directly affected bolus shape and tubing pressure. Although straight insertion is clinically preferred, our data suggest that an angled IIS elicits lower grades of ITR and delivers a bolus with lower tubing pressure and greater SA and V. The subcutaneous environment plays a crucial role in IIS longevity, and the insertion angle needs to be considered in future IIS designs and clinical trials.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.