Ambulatory blood pressure monitoring (ABPM) is being used increasingly in both clinical practice and hypertension research. Although there are many guidelines that emphasize the indications for ABPM, there is no comprehensive guideline dealing with all aspects of the technique. It was agreed at a consensus meeting on ABPM in Milan in 2011 that the 34 attendees should prepare a comprehensive position paper on the scientific evidence for ABPM.This position paper considers the historical background, the advantages and limitations of ABPM, the threshold levels for practice, and the cost-effectiveness of the technique. It examines the need for selecting an appropriate device, the accuracy of devices, the additional information and indices that ABPM devices may provide, and the software requirements.At a practical level, the paper details the requirements for using ABPM in clinical practice, editing considerations, the number of measurements required, and the circumstances, such as obesity and arrhythmias, when particular care needs to be taken when using ABPM.The clinical indications for ABPM, among which white-coat phenomena, masked hypertension, and nocturnal hypertension appear to be prominent, are outlined in detail along with special considerations that apply in certain clinical circumstances, such as childhood, the elderly and pregnancy, and in cardiovascular illness, examples being stroke and chronic renal disease, and the place of home measurement of blood pressure in relation to ABPM is appraised.The role of ABPM in research circumstances, such as pharmacological trials and in the prediction of outcome in epidemiological studies is examined and finally the implementation of ABPM in practice is considered in relation to the issue of reimbursement in different countries, the provision of the technique by primary care practices, hospital clinics and pharmacies, and the growing role of registries of ABPM in many countries.
In the last 30 years several organizations have developed protocols for clinical validation of blood pressure measuring devices. An international initiative was recently launched by the US Association for the Advancement of Medical Instrumentation (AAMI), the European Society of Hypertension Working Group on Blood Pressure Monitoring (ESH) and the International Organization for Standardization (ISO), aiming to reach consensus on a universal AAMI/ESH/ISO validation standard. The purpose of this statement by the ESH Working Group on Blood Pressure Monitoring is to provide practical guidance for investigators performing validation studies according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018), to ensure that its stipulations are meticulously implemented and data are fully reported. Thus, this statement provides: (i) a list of key recommendations for validation studies of intermittent non-invasive automated blood pressure measuring devices according to the AAMI/ESH/ISO Universal Standard, (ii) practical stepwise guidance for researchers performing these validation studies, (iii) a checklist for authors and reviewers of such studies.
Given the increasing use of ambulatory blood pressure monitoring (ABPM) in both clinical practice and hypertension research, a group of scientists, participating in the European Society of Hypertension Working Group on blood pressure monitoring and cardiovascular variability, in year 2013 published a comprehensive position paper dealing with all aspects of the technique, based on the available scientific evidence for ABPM. The present work represents an updated schematic summary of the most important aspects related to the use of ABPM in daily practice, and is aimed at providing recommendations for proper use of this technique in a clinical setting by both specialists and practicing physicians. The present article details the requirements and the methodological issues to be addressed for using ABPM in clinical practice, The clinical indications for ABPM suggested by the available studies, among which white-coat phenomena, masked hypertension, and nocturnal hypertension, are outlined in detail, and the place of home measurement of blood pressure in relation to ABPM is discussed. The role of ABPM in pharmacological, epidemiological, and clinical research is also briefly mentioned. Finally, the implementation of ABPM in practice is considered in relation to the situation of different countries with regard to the reimbursement and the availability of ABPM in primary care practices, hospital clinics, and pharmacies.
Use validated automated electronic upper-arm cu device.Prefer a device that takes triplicate readings automatically.If validated automated devices not available, use a manual electronic auscultatory device (LCD or LED display, or digital countdown, or good quality aneroid). De late at 2-3 mmHg/sec rate. Use 1st and 5th Korotko sound for systolic and diastolic BP.Annual maintenance of device is necessary.Select cu size according to the individual's arm circumference.Automated electronic devices: select cu size according to device instructions. Each electronic device has its own cu s, which are not interchangeable with those of other devices.Manual auscultatory devices: use a cu with bladder length at 75-100% of individual's arm circumference and width 37-50%. CUFF 2-3 o ice visits at 1-4-week intervals are usually required. At initial visit measure BP in both arms. Measure standing BP in treated hypertensives when there are symptoms suggesting postural hypotension.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.