Purpose To compare analytic and Monte Carlo-based algorithms for proton dose calculations in the lung, benchmarked against anthropomorphic lung phantom measurements. Methods A heterogeneous anthropomorphic moving lung phantom has been irradiated at numerous proton therapy centers. At five centers, the treatment plan could be calculated with both an analytic and Monte Carlo algorithm. The doses calculated in the treatment plans were compared with the doses delivered to the phantoms, which were measured using thermoluminescent dosimeters and film. Point doses were compared, as were planar doses using a gamma analysis. Results The analytic algorithms overestimated the dose to the center of the target by an average of 7.2%, whereas the Monte Carlo algorithms were within 1.6% of the physical measurements on average. In some regions of the target volume, the analytic algorithm calculations differed from the measurement by up to 31% in the iGTV (46% in the PTV), over-predicting the dose. All comparisons showed a region of at least 15% dose discrepancy within the iGTV between the analytic calculation and the measured dose. The Monte Carlo algorithm recalculations showed dramatically improved agreement with the measured doses, showing mean agreement within 4% for all cases, and a maximum difference of 12% within the iGTV. Conclusions Analytic algorithms often do a poor job predicting proton dose in lung tumors, overpredicting the dose to the target by up to 46%, and should not be used unless extensive validation counters the consistent results of the current study. Monte Carlo algorithms showed dramatically improved agreement with physical measurements and should be implemented to better reflect actual delivered dose distributions.
Purpose: Task Group (TG) 224 was established by the American Association of Physicists in Medicine's Science Council under the Radiation Therapy Committee and Work Group on Particle Beams. The group was charged with developing comprehensive quality assurance (QA) guidelines and recommendations for the three commonly employed proton therapy techniques for beam delivery: scattering, uniform scanning, and pencil beam scanning. This report supplements established QA guidelines for therapy machine performance for other widely used modalities, such as photons and electrons (TG 142, TG 40, TG 24, TG 22, TG 179, and Medical Physics Practice Guideline 2a) and shares their aims of ensuring the safe, accurate, and consistent delivery of radiation therapy dose distributions to patients. Methods: To provide a basis from which machine‐specific QA procedures can be developed, the report first describes the different delivery techniques and highlights the salient components of the related machine hardware. Depending on the particular machine hardware, certain procedures may be more or less important, and each institution should investigate its own situation. Results: In lieu of such investigations, this report identifies common beam parameters that are typically checked, along with the typical frequencies of those checks (daily, weekly, monthly, or annually). The rationale for choosing these checks and their frequencies is briefly described. Short descriptions of suggested tools and procedures for completing some of the periodic QA checks are also presented. Conclusion: Recommended tolerance limits for each of the recommended QA checks are tabulated, and are based on the literature and on consensus data from the clinical proton experience of the task group members. We hope that this and other reports will serve as a reference for clinical physicists wishing either to establish a proton therapy QA program or to evaluate an existing one.
Purpose: To analyze the most recent results of the Imaging and Radiation Oncology Core Houston Quality Assurance Center's (IROC-H) anthropomorphic head and neck (H&N) phantom to determine the nature of failing irradiations and the feasibility of altering credentialing criteria. Methods: IROC-H's H&N phantom, used for intensity-modulated radiation therapy credentialing for National Cancer Institute-sponsored clinical trials, requires that an institution's treatment plan agrees within ±7% of measured thermoluminescent dosimeter (TLD) doses; it also requires that ≥85% of pixels pass ±4 mm distance to agreement (7%/4 mm gamma analysis for film). The authors re-evaluated 156 phantom irradiations (November 1, 2014-October 31, 2015 according to the following tighter criteria: (1) 5% TLD and 5%/4 mm, (2) 5% TLD and 5%/3 mm, (3) 4% TLD and 4%/4 mm, and (4) 3% TLD and 3%/3 mm. Failure rates were evaluated with respect to individual film and TLD performance by location in the phantom. Overall poor phantom results were characterized qualitatively as systematic errors (correct shape and position but wrong magnitude of dose), setup errors/positional shifts, global but nonsystematic errors, and errors affecting only a local region. Results: The pass rate for these phantoms using current criteria was 90%. Substituting criteria 1-4 reduced the overall pass rate to 77%, 70%, 63%, and 37%, respectively. Statistical analyses indicated that the probability of noise-induced TLD failure, even at the 5% criterion, was <0.5%. Phantom failures were generally identified by TLD (≥66% failed TLD, whereas ≥55% failed film), with most failures occurring in the primary planning target volume (≥77% of cases). Results failing current criteria or criteria 1 were primarily diagnosed as systematic >58% of the time (11/16 and 21/36 cases, respectively), with a greater extent due to underdosing. Setup/positioning errors were seen in 11%-13% of all failing cases (2/16 and 4/36 cases, respectively). Local errors (8/36 cases) could only be demonstrated at criteria 1. Only three cases of global errors were identified in these analyses. For current criteria and criteria 1, irradiations that failed from film only were overwhelmingly associated with phantom shifts/setup errors (≥80% of cases). Conclusions: This study highlighted that the majority of phantom failures are the result of systematic dosimetric discrepancies between the treatment planning system and the delivered dose. Further work is necessary to diagnose and resolve such dosimetric inaccuracy. In addition, the authors found that 5% TLD and 5%/4 mm gamma criteria may be both practically and theoretically achievable as an alternative to current criteria. C
Purpose: With external beam radiation therapy, uncertainties in treatment planning and delivery can result in an undesirable dose distribution delivered to the patient that can compromise the benefit of treatment. Techniques including geometric margins and probabilistic optimization have been used effectively to mitigate the effects of uncertainties. However, their broad application is inconsistent and can compromise the conclusions derived from cross-technique and cross-modality comparisons. Methods and Materials: Conventional methods to deal with treatment planning and delivery uncertainties are described, and robustness analysis is presented as a framework that is applicable across treatment techniques and modalities. Results: This report identifies elements that are imperative to include when conducting a robustness analysis and describing uncertainties and their dosimetric effects.NotedEarn CME credit by taking a brief online assessment at https://academy.astro.org.Conclusion: The robustness analysis approach described here is presented to promote reliable plan evaluation and dose reporting, particularly during clinical trials conducted across institutions and treatment modalities.
Purpose: Accurate data regarding linear accelerator (Linac) radiation characteristics are important for treatment planning system modeling as well as regular quality assurance of the machine. The Imaging and Radiation Oncology Core-Houston (IROC-H) has measured the dosimetric characteristics of numerous machines through their on-site dosimetry review protocols. Photon data are presented and can be used as a secondary check of acquired values, as a means to verify commissioning a new machine, or in preparation for an IROC-H site visit. Methods: Photon data from IROC-H on-site reviews from 2000 to 2014 were compiled and analyzed. Specifically, data from approximately 500 Varian machines were analyzed. Each dataset consisted of point measurements of several dosimetric parameters at various locations in a water phantom to assess the percentage depth dose, jaw output factors, multileaf collimator small field output factors, off-axis factors, and wedge factors. The data were analyzed by energy and parameter, with similarly performing machine models being assimilated into classes. Common statistical metrics are presented for each machine class. Measurement data were compared against other reference data where applicable. Results: Distributions of the parameter data were shown to be robust and derive from a student's t distribution. Based on statistical and clinical criteria, all machine models were able to be classified into two or three classes for each energy, except for 6 MV for which there were eight classes. Quantitative analysis of the measurements for 6, 10, 15, and 18 MV photon beams is presented for each parameter; supplementary material has also been made available which contains further statistical information.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.