In this work the extent of the linear dose response and the dynamic dose range of N-vinylpyrrolidone-argon based (VIPAR) polymer gels were investigated. VIPAR gels were irradiated using a 6 MV linear accelerator up to 60 Gy and a Nucletron microSelectron 192Ir HDR brachytherapy source to much higher doses to cover a dose range of two orders of magnitude. They were then MR scanned at 1.5 T to obtain T2-maps. VIPAR gel measurements obtained from the two irradiation regimes were calibrated against ion chamber measurements and dose calculations derived using the AAPM TG-43 protocol respectively. A satisfying agreement between the calibration results derived using the 6 MV x-rays and the 192Ir source was found for doses up to 60 Gy, implying that the response of the VIPAR gels is independent of photon energy and dose rate. A linear R2 dose response up to approximately 40 Gy and a dynamic dose range up to at least approximately 250 Gy were observed. VIPAR gel dose measurements derived using the monoexponentially fitted brachytherapy calibration data were found to be quite accurate.
In this work, the utilization of polymer gel-MRI dosimetry for measurements at distances relevant to clinical brachytherapy and intravascular applications [i.e., in the mm range, where steep three-dimensional (3-D) dose gradients exist] is investigated using N-vinylpyrrolidone-based gels. Transverse axis radial dose distributions, dose distributions parallel to the source axis, and 2-D dose distributions around the commonly used microSelectron 192Ir HDR source are measured for single source dwell position irradiations. Experimental results are found in good agreement with verified Monte Carlo calculations, even for distances less than 3 mm from the source. The effect of various MRI parameters, such as slice thickness, slice mispositioning, and in-plane resolution, on the accuracy of the method is also investigated. Possible limitations of the method are discussed, and its' overall potential in brachytherapy dosimetry is evaluated. Experimental 2-D dose distributions for an intravascular application following the Paris irradiation protocol are compared to corresponding commercial treatment planning system calculations. Results suggest that polymer gel-MRI dosimetry is capable of experimentally verifying dose distributions in relevant clinical intravascular applications.
VIPAR polymer gels and 3D MRI techniques were evaluated for their ability to provide experimental verification of 3D dose distributions in a simulation of a 192Ir prostate monotherapy clinical application. A real clinical treatment plan was utilized, generated by post-irradiation, CT based calculations derived from Plato BPS and Swift treatment planning systems. The simulated treatment plan involved the use of 10 catheters and 39 source positions within a glass vessel of appropriate dimensions, homogeneously filled with the VIPAR gel. 3D high resolution MR scanning of the gel produced T2 relaxation time maps, from which 3D dose distributions were derived via an appropriate calibration procedure. Results were compared to corresponding dose distributions obtained from the Plato and Swift treatment planning systems. Quantitative comparison, on a point by point basis, was based on user adopted acceptance criteria of 5% dose-difference and 3 mm distance-to-agreement. Significant deviations between experimental and calculated dose distributions were found for doses lower than 50% due to the reduced dose resolution of the method in the low dose, low dose gradient region. Measurement errors were observed at 1.0-1.5 mm around each catheter due to MR imaging susceptibility artifacts. For most remaining points the acceptance criteria were fulfilled. Systematic offsets of the order of 1-2 mm, observed between measured and corresponding calculated isocontours at specific segments, are attributed to the 1 mm uncertainty in catheter reconstruction and 1 mm uncertainty in the alignment of the MR and CT imaging planes.
In this work, three-dimensional (3-D) MRI techniques are employed in N-Vinylpyrrolidone-Argon-(VIPAR-) based polymer gel dosimetry. VIPAR gels were irradiated using a Nucletron microSelection 192Ir HDR brachytherapy remote afterloading system with single source dwell position and intravascular brachytherapy irradiation protocols. A single VIPAR gel and a single irradiation are adequate to obtain the full calibration curve needed. The 3-D dose distributions obtained with the 3-D MRI method were found to be in good agreement with the corresponding Monte Carlo calculations, for brachytherapy and intravascular irradiations. The method allows the reconstruction of isodose contours over any plane, with increased spatial resolution and accuracy following a single MR acquisition. VIPAR gel measurements using a 3-D MRI readout technique can be of particular use in the experimental dosimetry of brachytherapy sources, as well as for dose verification purposes when complex irradiation regimes and three-dimensional dose gradients are investigated.
Background/Aims: There is no treatment, without side effects, efficiently preventing or curing skin burns, caused by radiotherapy. A new experimental topical treatment protocol was assessed in mice receiving orthovoltage X-rays at an equivalent dose to that applied to human breast cancer patients in conventional radiotherapy. Methods: SKH-HR2 female hairless mice were irradiated on their dorsum with a total dose of 4,300 cGy during a 1-month period (20 fractions). The treatment group received a combination of 3 topical products, an oil-in-water cream, a gel containing Pinus halepensis bark aqueous extract, and an ointment containing olive oil extract of the marine isopod Ceratothoa oestroides. The positive control group was treated with a conventionally used commercial gel, whereas the negative control group did not receive any topical treatment. Skin alterations were evaluated by macroscopic examinations, measurements of transepidermal water loss (TEWL), melanin content, erythema intensity, hydration, and histopathology assessment. Results: Sixty days after radiation, TEWL and hydration values were abnormal and elements of acute, chronic, and granulomatous inflammation were present in all cases. The severest damage was detected in the deeper dermis. Treatment showed a comparatively beneficial effect on chronic and granulomatous inflammation while positive control was beneficial on acute inflammation. Conclusion: Skin anti-inflammatory treatment was the most effective but must be applied for several months. Further preclinical studies should be conducted, assimilating a human cancer radiation therapeutic schema with the aim of optimizing skin inflammation treatment.
The aim of this investigation was the evaluation of the usefulness of N-vinyl pyrrolidone argon (VIPAR) polymer gel dosimetry for relative dose measurements using the majority of types and energies of radiation beams used in clinical practice. For this reason, VIPAR polymer gels were irradiated with the following beams: 6 and 23 MV photons (maximum dose: 15 Gy) and 6, 9, 12, 15, 18 and 21 MeV electrons (90% dose: 15 Gy). Using 6 MV x-rays, a linear gel dose response was verified for doses up to 20 Gy. Assuming linearity of response for the rest of the photon and electron beams used in this study, percentage depth dose measurements were derived. For all beams used and the range of relative doses studied, a satisfying agreement was observed between percentage depth dose measurements performed using the VIPAR gel-MRI method and an ion chamber, validating the assumption that a linear gel dose response holds for all photon and electron beams studied. VIPAR gels, therefore, can be used for relative dose distribution measurements using photons or electrons of any typical energy used in external radiotherapy applications. It is also demonstrated that two-dimensional dose distribution measurements through an irradiated (9 MeV electrons, 3 cm x 3 cm cone) VIPAR gel volume can be easily obtained.
Greek Atomic Energy Commission's Department of Licensing and Inspections conducted a national survey for the establishment of nuclear medicine (NM) dose reference levels (DRLs) for adult patients, in Greece. The administered activities (AAs) (MBq) were collected from 120 NM departments (88 % of total), during on-site inspections for licensing purposes. Factors influencing the image quality were also investigated. The established national DRLs represent the AA value corresponding to the 75th percentile of the AA frequency distributions. In their majority, national DRLs and average AAs are comparable with the ones published in the international literature. In the light of new technologies, there might be potential for reducing the higher values of AAs, in co-operation with the nuclear medicine experts.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.